Channel discrimination along all contacts of the cochlear implant electrode array and its relation to speech perception
Open Access
- 15 February 2019
- journal article
- research article
- Published by Taylor & Francis in International Journal of Audiology
- Vol. 58 (5) , 262-268
Abstract
Objective: To test the channel discrimination of cochlear implant (CI) users along all contacts of the electrode array and assess whether this is related to speech perception. Design: CI recipients were tested with a custom-made channel discrimination test. They were asked to distinguish a target stimulus from two reference stimuli in a three-alternative forced choice (3AFC) task. The target stimulus was evoked using current steering, with current steering coefficients (α) of 1, 0.5 and 0.25. The test provided a discrimination score (Dα) for each electrode contact along the array. Study sample: Thirty adults implanted with a CI from Advanced Bionics. Results: Large variations in Dα scores were observed, both across the electrode array and between subjects. Statistical analysis revealed a significant channel-to-channel variability in Dα score (p < 0.01). Further, there was a significant relationship between subjects’ Dα scores and their speech perception in quiet (p < 0.001). Conclusions: The large variations in Dα score emphasise the importance of testing pitch discrimination across the complete electrode array. The relationship between Dα score and speech perception indicates that pitch discrimination might be a contributing factor to the performance of individual implant users.Keywords
Funding Information
- Stichting voor de Technische Wetenschappen
This publication has 28 references indexed in Scilit:
- Spread of Excitation and Channel Interaction in Single- and Dual-Electrode Cochlear Implant StimulationEar & Hearing, 2012
- A SIMPLE ALTERNATIVE ANALYSIS FOR THRESHOLD DATA DETERMINED BY ASCENDING FORCED‐CHOICE METHODS OF LIMITSJournal of Sensory Studies, 2010
- Simultaneous and non-simultaneous dual electrode stimulation in cochlear implants: evidence for two neural response modalitiesActa Oto-Laryngologica, 2009
- Relation Between Neural Response Telemetry Thresholds, T- and C-Levels, and Loudness Judgments in 12 Adult Nucleus 24 Cochlear Implant RecipientsEar & Hearing, 2007
- Using Current Steering to Increase Spectral Resolution in CII and HiRes 90K UsersEar & Hearing, 2007
- Across-Site Threshold Variation in Cochlear Implants: Relation to Speech RecognitionAudiology and Neurotology, 2004
- Place and temporal cues in pitch perception: are they truly independent?Acoustics Research Letters Online, 2000
- Comparison of electrode discrimination, pitch ranking, and pitch scaling data in postlingually deafened adult cochlear implant subjectsThe Journal of the Acoustical Society of America, 1997
- Electrode ranking of ‘‘place pitch’’ and speech recognition in electrical hearingThe Journal of the Acoustical Society of America, 1995
- Pitch perception by cochlear implant subjectsThe Journal of the Acoustical Society of America, 1987