An in vitro study of the reduction in wear of metal-on-metal hip prostheses using surface-engineered femoral heads
Open Access
- 1 April 2002
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 216 (4) , 219-230
Abstract
Although the wear of existing metal-on-metal (MOM) hip prostheses (1 mm3/106 cycles) is much lower than the more widely used polyethylene-on-metal bearings, there are concerns about the toxicity of metal wear particles and elevated metal ion levels, both locally and systemically, in the human body. The aim of this study was to investigate the possibility of reducing the volume of wear, the concentration of metal debris and the level of metal ion release through using surfaceengineered femoral heads. Three thick (8-12 μm) coatings (TiN, CrN and CrCN) and one thin (2 μm) coating (diamond-like carbon, DLC), were evaluated on the femoral heads when articulating against high carbon content cobalt-chromium alloy acetabular inserts (HC CoCrMo) and compared with a clinically used MOM cobalt-chromium alloy bearing couple using a physiological anatomical hip joint simulator (Leeds Mark II). This study showed that CrN, CrCN and DLC coatings produced substantially lower wear volumes for both the coated femoral heads and the HC CoCrMo inserts. The TiN coating itself had little wear, but it caused relatively high wear of the HC CoCrMo inserts compared with the other coatings. The majority of the wear debris for all half-coated couples comprised small, 30 nm or less, CoCrMo metal particles. The Co, Cr and Mo ion concentrations released from the bearing couples of CrN-, CrCN- and DLC-coated heads articulating against HC CoCrMo inserts were at least 7 times lower than those released from the clinical MOM prostheses. These surface-engineered femoral heads articulating on HC CoCrMo acetabular inserts produced significantly lower wear volumes and rates, and hence lower volumetric concentrations of wear particles, compared with the clinical MOM prosthesis. The substantially lower ion concentration released by these surface-engineered components provides important evidence to support the clinical application of this technology.Keywords
This publication has 16 references indexed in Scilit:
- Microhardness and structural analysis of (Ti,Al)N, (Ti,Cr)N, (Ti,Zr)N and (Ti,V)N filmsJournal of Vacuum Science & Technology A, 2000
- Comparison of mechanical properties of TiN thin films using nanoindentation and bulge testThin Solid Films, 1998
- TiN coating wear mechanisms in dry sliding contact against high speed steelSurface and Coatings Technology, 1998
- TiN hard coating prepared by sputter ion plating system with facing target sputtering source and RF dischargeThin Solid Films, 1998
- Mechanical stability of DLC films on metallic substrates: Part I—Film structure and residual stress levelsThin Solid Films, 1998
- Cobalt and Chromium Concentrations in Patients With Metal on Metal Total Hip ReplacementsClinical Orthopaedics and Related Research, 1996
- In Vivo Wear of 3 Types of Metal on Metal Hip Prostheses During 2 Decades of UseClinical Orthopaedics and Related Research, 1996
- Tissue Reaction to Metal on Metal Total Hip ProsthesesClinical Orthopaedics and Related Research, 1996
- The differences in toxicity and release of bone-resorbing mediators induced by titanium and cobalt-chromium-alloy wear particles.Journal of Bone and Joint Surgery, 1993
- Size of metallic and polyethylene debris particles in failed cemented total hip replacementsThe Journal of Bone and Joint Surgery. British volume, 1992