Flat response detectors for the vacuum ultraviolet and soft x-ray region: InP:Fe photoconductors
Authors:
- 1 October 1986
- journal article
- conference paper
- Published by AIP Publishing in Journal of Applied Physics
- Vol. 60 (7) , 2596-2599
Abstract
We have demonstrated that InP:Fe photoconductive detectors are sensitive to radiation between 20 and 3000 eV. Within experimental error the measured response is flat in this energy range, in agreement with theory. This property, coupled with the simplicity and high speed (full width at one half maximum of less than 200 ps) of these devices, makes them very interesting for the measurement of pulsed vacuum ultraviolet and x-ray radiation and the total instantaneous integrated radiative power emitted from a pulsed source. Low fidelity response can occur in these devices for excitation times that are many times longer than the primary response time of the detector.Keywords
This publication has 6 references indexed in Scilit:
- Picosecond Photoconductors as Radiation DetectorsIEEE Transactions on Nuclear Science, 1986
- Pulsed soft x-ray response of InP:Fe photoconductorsApplied Physics Letters, 1984
- Picosecond photoresponse in 3He+ bombarded InP photoconductorsApplied Physics Letters, 1984
- Use of a laser-produced plasma as a source of focused vacuum-ultraviolet continuum radiationOptics Letters, 1979
- Investigation of minority-carrier diffusion lengths by electron bombardment of Schottky barriersJournal of Applied Physics, 1978
- Statistics of the Recombinations of Holes and ElectronsPhysical Review B, 1952