Correlation of Contrast-Detail Analysis and Clinical Image Quality Assessment in Chest Radiography with a Human Cadaver Study
Authors:
- An De CropACAn De Crop
- Klaus BacherKBKlaus Bacher
- Tom Van HoofTHTom Van Hoof
- Peter V. SmeetsPSPeter V. Smeets
- Barbara S. SmetBSBarbara S. Smet
- Merel VergauwenMVMerel Vergauwen
- Urszula KiendysUKUrszula Kiendys
- Philippe DuyckPDPhilippe Duyck
- Koenraad VerstraeteKVKoenraad Verstraete
- Katharina D’hErdeKDKatharina D’hErde
- Hubert ThierensHTHubert Thierens
- 1 January 2012
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 262 (1) , 298-304
Abstract
To determine the correlation between the clinical and physical image quality of chest images by using cadavers embalmed with the Thiel technique and a contrast-detail phantom. The use of human cadavers fulfilled the requirements of the institutional ethics committee. Clinical image quality was assessed by using three human cadavers embalmed with the Thiel technique, which results in excellent preservation of the flexibility and plasticity of organs and tissues. As a result, lungs can be inflated during image acquisition to simulate the pulmonary anatomy seen on a chest radiograph. Both contrast-detail phantom images and chest images of the Thiel-embalmed bodies were acquired with an amorphous silicon flat-panel detector. Tube voltage (70, 81, 90, 100, 113, 125 kVp), copper filtration (0.1, 0.2, 0.3 mm Cu), and exposure settings (200, 280, 400, 560, 800 speed class) were altered to simulate different quality levels. Four experienced radiologists assessed the image quality by using a visual grading analysis (VGA) technique based on European Quality Criteria for Chest Radiology. The phantom images were scored manually and automatically with use of dedicated software, both resulting in an inverse image quality figure (IQF). Spearman rank correlations between inverse IQFs and VGA scores were calculated. A statistically significant correlation (r = 0.80, P < .01) was observed between the VGA scores and the manually obtained inverse IQFs. Comparison of the VGA scores and the automated evaluated phantom images showed an even better correlation (r = 0.92, P < .001). The results support the value of contrast-detail phantom analysis for evaluating clinical image quality in chest radiography.Keywords
This publication has 22 references indexed in Scilit:
- Dose and perceived image quality in chest radiographyEuropean Journal of Radiology, 2009
- Contrast-detail evaluation and dose assessment of eight digital chest radiography systems in clinical practiceEuropean Radiology, 2005
- Assessment and optimisation of the image quality of chest-radiography systemsRadiation Protection Dosimetry, 2005
- Dose Reduction in Patients Undergoing Chest Imaging: Digital Amorphous Silicon Flat-Panel Detector Radiography Versus Conventional Film-Screen Radiography and Phosphor-Based Computed RadiographyAmerican Journal of Roentgenology, 2003
- An experimental comparison of detector performance for direct and indirect digital radiography systemsMedical Physics, 2003
- Evaluation of the imaging properties of two generations of a CCD‐based system for digital chest radiographyMedical Physics, 2002
- The influence of different technique factors on image quality of chest radiographs as evaluated by modified CEC image quality criteriaBritish Journal of Radiology, 2002
- Comparison of an amorphous silicon/cesium iodide flat‐panel digital chest radiography system with screen/film and computed radiography systems — A contrast‐detail phantom studyMedical Physics, 2001
- Image quality vs radiation dose for a flat-panel amorphous silicon detector: a phantom studyEuropean Radiology, 2001
- Methods for the Evaluation of Image Quality: A ReviewRadiation Protection Dosimetry, 2000