A new planar multielectrode array: recording from a rat auditory cortex
Authors:
- Vassiliy TsytsarevVTVassiliy Tsytsarev
- Makoto TaketaniMTMakoto Taketani
- Frank SchottlerFSFrank Schottler
- Shigeru TanakaSTShigeru Tanaka
- Masahiko HaraMHMasahiko Hara
- 6 October 2006
- journal article
- Published by IOP Publishing in Journal of Neural Engineering
- Vol. 3 (4) , 293-298
Abstract
In the study of the spatiotemporal properties of the cortex, a demand often arises for recording local field evoked potentials (LFEP) and neural spikes from a quantity of points at close range from each other. In such a situation a device composed of a lot of electrodes assembled in a single bunch would be suitable. Such circumstances gave us the impetus to create the device described in this paper, namely a new planar electrode array for in vivo multisite extracellular recording. The device is made of plastic and includes platinum electrodes 50 microm in diameter. The array of 64 incorporated microelectrodes is placed on the surface of the cortex of anesthetized rats. Recordings could be made through all electrodes for more than 1 h without damage to the cortex. The inter-polar distance is approximately 100 microm so that each individual electrode can record activity from a separate population of neurons near the cortical surface. The recording system described here is highly useful for visualizing spatiotemporal structure of the cortical activities and for imaging dynamic neuronal assemblies.Keywords
This publication has 20 references indexed in Scilit:
- Chronic, multisite, multielectrode recordings in macaque monkeysProceedings of the National Academy of Sciences, 2003
- Independently movable multielectrode array to record multiple fast-spiking neurons in the cerebral cortex during cognitionMethods, 2003
- Multielectrode recordings: the next stepsCurrent Opinion in Neurobiology, 2002
- Redundancy and Synergy Arising from Pairwise Correlations in Neuronal EnsemblesJournal of Computational Neuroscience, 2002
- Parallel sensingNature, 2000
- Topographic Analysis of Epidural Pure-Tone–Evoked Potentials in Gerbil Auditory CortexJournal of Neurophysiology, 2000
- Three-dimensional spatial selectivity of hippocampal neurons during space flightNature Neuroscience, 2000
- A multielectrode implant device for the cerebral cortexJournal of Neuroscience Methods, 1999
- A new planar multielectrode array for extracellular recording: application to hippocampal acute sliceJournal of Neuroscience Methods, 1999
- Reactivation of Hippocampal Cell Assemblies: Effects of Behavioral State, Experience, and EEG DynamicsThe Journal of Neuroscience, 1999