Development and calibration of a multi-delay coherence imaging diagnostic on the MAST-U tokamak
Open Access
- 1 May 2024
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 95 (5)
Abstract
The MAST-U Super-X divertor provides the opportunity to study fusion plasma exhaust under novel conditions. However, in order to study these conditions, advanced diagnostics are required. Following the development of the MAST-U Multi-Wavelength Imaging (MWI) diagnostic, we present the installation of a multi-delay coherence imaging spectroscopy (CIS) system within the MAST-U MWI, along with modifications made to the MWI for effective operation. This diagnostic will measure either carbon ion flow velocities and temperatures or electron densities through Dγ emission. We have extended previously developed techniques for wavelength calibration to account for errors due to the misalignment of interferometer components. In addition, we have developed a comprehensive calibration procedure to account for the temperature dependence of the instrument’s delays by fitting to a linearly modified version of the delay equation presented by Veiras et al. [Appl. Opt. 49(15), 2769 (2010)]. Together, these procedures reduce the cost and hardware complexity of implementing CIS instruments when compared to those that use in situ or tunable laser calibration systems, as calibrations can be generated to good accuracy using previously measured data.Keywords
Funding Information
- EUROfusion (10105220)
- Engineering and Physical Sciences Research Council (EP/W006839/1)
This publication has 21 references indexed in Scilit:
- Publisher’s Note: “Coherence imaging spectroscopy at Wendelstein 7-X for impurity flow measurements” [Rev. Sci. Instrum. 91, 013501 (2020)]Review of Scientific Instruments, 2020
- MANTIS: A real-time quantitative multispectral imaging system for fusion plasmasReview of Scientific Instruments, 2019
- A new calibration implementation for Doppler Coherence Imaging SpectroscopyFusion Engineering and Design, 2019
- Calibrated helium and carbon ion flow measurements in the DIII-D divertor plasmaNuclear Materials and Energy, 2019
- Laser calibration of the DIII-D coherence imaging systemReview of Scientific Instruments, 2018
- Coherence imaging of scrape-off-layer and divertor impurity flows in the Mega Amp Spherical Tokamak (invited)Review of Scientific Instruments, 2014
- C-III flow measurements with a coherence imaging spectrometerReview of Scientific Instruments, 2012
- Super-X divertors and high power density fusion devicesPhysics of Plasmas, 2009
- High-speed high-resolution plasma spectroscopy using spatial-multiplex coherence imaging techniques (invited)Review of Scientific Instruments, 2006
- Measurements and modeling of ion and neutral distribution functions in a partially ionized magnetically confined argon plasmaPhysics of Plasmas, 2004