Comments on ‘The effective depth of cylindrical ionization chambers in water for clinical proton beams’
Author:
- 12 October 2012
- journal article
- editorial
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 57 (21) , 7219-7224
Abstract
In a recent paper (Bhullar and Watchman 2012 Phys. Med. Biol. 57 273-86), the authors comment on data concerning the effective depth of cylindrical ionization chambers in water for clinical proton beams calculated in a previous paper (Palmans 2006 Phys. Med. Biol. 51 3483-501) and propose different results. They present a closed-format expression for an integral in the analytical model and claim that the series expansion I used in the older paper does not converge to the correct solution. They also claim that this is the reason for the resulting values of the shift of the effective depth with respect to the chamber centre to be different in both papers. Both claims are, however, incorrect as I show in this comment. The values they present are most likely based on a mistake in the scaling coefficients to account for the non-water equivalence of wall, sleeve and central electrode materials. The better agreement which they observe between their values and the recommendation of IAEA TRS-398 is thus a coincidence. My conclusion is that the best available data for the effective depth of cylindrical ionization chambers in water in clinical proton beams are at present the values I published before (Palmans 2006 Phys. Med. Biol. 51 3483-501) and which are supported by Monte Carlo simulations and experimental evidence.Keywords
This publication has 10 references indexed in Scilit:
- The effective depth of cylindrical ionization chambers in water for clinical proton beamsPhysics in Medicine & Biology, 2011
- Perturbation factors for cylindrical ionization chambers in proton beams. Part I: corrections for gradientsPhysics in Medicine & Biology, 2006
- A small-body portable graphite calorimeter for dosimetry in low-energy clinical proton beamsPhysics in Medicine & Biology, 2004
- Cross-calibration of ionization chambers in proton and carbon beamsPhysics in Medicine & Biology, 2004
- Experimentalpwallandpcelcorrection factors for ionization chambers in low-energy clinical proton beamsPhysics in Medicine & Biology, 2001
- Depth ionization curves for an unmodulated proton beam measured with different ionization chambersMedical Physics, 2000
- On the effective point of measurement of cylindrical ionization chambers for proton beams and other heavy charged particle beamsPhysics in Medicine & Biology, 2000
- Effective point of measurement of cylindrical ionization chambers for heavy charged particlesPhysics in Medicine & Biology, 2000
- Monte Carlo study of fluence perturbation effects on cavity dose response in clinical proton beamsPhysics in Medicine & Biology, 1998
- Calculated depth dose distributions for proton beams in some low-ZmaterialsPhysics in Medicine & Biology, 1997