Effects of flow rate on high‐throughput quantitative analysis of protein‐precipitated plasma using liquid chromatography/tandem mass spectrometry
- 4 February 2002
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 16 (6) , 537-543
Abstract
The effects of flow rate and column length on analyte response (peak area and height), total cycle time, column backpressure, and elution volume are presented. Rapid chromatographic separations and tandem mass spectrometric (MS/MS) detection are applied to the supernatant of protein‐precipitated plasma standards containing four compounds from a drug discovery screen. The plasma samples were injected onto three C‐18 columns (2 × 10,2.1 × 30 and 2.1 × 50 mm) at flow rates of 0.25, 0.50, 1.00 and 1.50 mL/min. The plasma samples were detected using a Sciex API 3000 tandem mass spectrometer operated in the Turbo Ionspray mode. A post‐column split was used to maintain a flow rate of 0.25 mL/min into the mass spectrometer source to avoid differences in nebulization efficiency. The data show that diluted protein‐precipitated plasma supernatants show average matrix effects (i.e. suppression) of 60.0% (2 × 10 mm), 89.3% (2 × 30 mm), and 76.7% (2 × 50 mm) of expected response at 10 ng/mL. Average matrix effects of 70.2% (2 × 10 mm), 88.9% (2 × 30 mm), and 81.2% (2 × 50 mm) of expected response at 1000 ng/mL plasma. The data also show if peak widths remain relatively constant, analytes are less sensitive as flow rates are increased. These data are consistent with the concentration‐dependent relationship of ionspray in the range of flow rates studied. The data show that, while analyte response decreased proportionately to increases in flow rate, the analysis cycle times did not decrease proportionately. Copyright © 2002 John Wiley & Sons, Ltd.Keywords
This publication has 15 references indexed in Scilit:
- High‐speed liquid chromatography/tandem mass spectrometry using a monolithic column for high‐throughput bioanalysisRapid Communications in Mass Spectrometry, 2001
- Ultra‐fast gradient vs. fast isocratic chromatography in bioanalytical quantification by liquid chromatography/tandem mass spectrometryRapid Communications in Mass Spectrometry, 2001
- Ultrafast liquid chromatography/ultraviolet and liquid chromatography/tandem mass spectrometric analysisRapid Communications in Mass Spectrometry, 2000
- Mechanistic investigation of ionization suppression in electrospray ionizationJournal of the American Society for Mass Spectrometry, 2000
- Parallel ultra-high flow rate liquid chromatography with mass spectrometric detection using a multiplex electrospray source for direct, sensitive determination of pharmaceuticals in plasma at extremely high throughputRapid Communications in Mass Spectrometry, 2000
- High-throughput quantitative bioanalysis by LC/MS/MSBiomedical Chromatography, 2000
- Bioanalytical applications of ?fast chromatography? to high-throughput liquid chromatography/tandem mass spectrometric quantitationRapid Communications in Mass Spectrometry, 2000
- Ion spray mass spectrometric detection for liquid chromatography: A concentration- or a mass-flow-sensitive device?Journal of Chromatography A, 1993
- Orientational dynamics of a hydrophobic guest in a chromatographic stationary phase: effect of wetting by alcoholAnalytical Chemistry, 1992
- Reduction of reequilibration time following gradient elution reversed-phase liquid chromatographyAnalytical Chemistry, 1990