Improving Sensitivity of Arterial Spin Labeling Perfusion MRI in Alzheimer's Disease Using Transfer Learning of Deep Learning‐Based ASL Denoising
Authors:
- Lei ZhangLZLei Zhang
- Aldo CamargoACAldo Camargo
- Ze WangZe Wang
- Alzheimer'S Disease Neuroimaging InitiativeAIAlzheimer'S Disease Neuroimaging Initiative
- 6 November 2021
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 55 (6) , 1710-1722
Abstract
Background: Arterial spin labeling (ASL) perfusion magnetic resonance imaging (MRI) denoising through deep learning (DL) often faces insufficient training data from patients. One solution is to train DL models using healthy subjects' data which are more widely available and transfer them to patients' data.Purpose: To evaluate the transferability of a DL‐based ASL MRI denoising method (DLASL).Study Type: Retrospective.Subjects: Four hundred and twenty‐eight subjects (189 females) from three cohorts.Field Strength/Sequence: 3 T two‐dimensional (2D) echo‐planar imaging (EPI)‐based pseudo‐continuous ASL (PCASL) and 2D EPI‐based pulsed ASL (PASL) sequences.Assessment: DLASL was trained using young healthy adults' PCASL data (Dataset 1: 250/30 subjects as training/validation set) and was directly transferred (DTF) to PCASL data from Dataset 2 (45 subjects test set) of normal controls (NC) and Alzheimer's disease (AD) groups. DLASL was fine‐tuned (DLASLFT) and tested on PASL data from Dataset 3 (103 subjects test set) of NC and AD. An existing non‐DL method (NonDL) was used for comparison. Cerebral blood flow (CBF) images from ASL MRI were compared between NC and AD to assess characteristic hypoperfusion (lower CBF) patterns in AD. CBF image quality and CBF map sensitivity for detecting hypoperfusion using peak t‐value and suprathreshold cluster size are outcome measures.Statistical Tests: Paired t‐test, two‐sample t‐test, one‐way analysis of variance, and Tukey honestly significant difference, and linear mixed‐effects models were used. P < 0.05 was considered statistically significant.Results: Mean contrast‐to‐noise ratio (CNR) of Dataset 2 showed that DTF outperformed NonDL (AD: 3.38 vs. 2.64, NC: 3.80 vs. 3.36). On Dataset 3, DLASLFT outperformed NonDL measured by mean CNR (AD: 2.45 vs. 1.87, NC: 2.54 vs. 2.17) and mean radiologic score (2.86 vs. 2.44). Image quality improvement was significant on both test sets. DTF and DLASLFT improved sensitivity for detecting AD‐related hypoperfusion patterns compared with NonDL.Data Conclusion: We demonstrated the DLASL's transferability across different ASL sequences and different populations.Level of Evidence: 3Technical Efficacy: Stage 2Keywords
Funding Information
- AbbVie
- Alzheimer's Association
- Alzheimer's Drug Discovery Foundation
- BioClinica
- Biogen
- Canadian Institutes of Health Research
- U.S. Department of War (W81XWH‐12‐2‐0012)
- Eisai Incorporated
- Eli Lilly and Company
- Roche
- Foundation for the National Institutes of Health
- GE Healthcare
- Genentech
- H. Lundbeck A/S
- Merck
- National Institute of Biomedical Imaging and Bioengineering
- National Institute on Aging
- National Institutes of Health
- National Institute on Aging (R01AG060054)
- National Institutes of Health (R01 AG070227, R01EB031080‐01A1, P41EB029460‐01A1)
- Northern California Institute for Research and Education
- Novartis Pharmaceuticals Corporation
- Pfizer
- Servier
- Takeda Pharmaceutical Company
- Fujirebio US
- Norman Cousins Center for Psychoneuroimmunology
This publication has 29 references indexed in Scilit:
- On the Definition of Signal-To-Noise Ratio and Contrast-To-Noise Ratio for fMRI DataPLOS ONE, 2013
- Improving cerebral blood flow quantification for arterial spin labeled perfusion MRI by removing residual motion artifacts and global signal fluctuationsMagnetic Resonance Imaging, 2012
- Applications of arterial spin labeled MRI in the brainJournal of Magnetic Resonance Imaging, 2012
- Test–retest reliability of arterial spin labeling with common labeling strategiesJournal of Magnetic Resonance Imaging, 2011
- Unbiased average age-appropriate atlases for pediatric studiesNeuroImage, 2011
- Distinct cerebral perfusion patterns in FTLD and ADNeurology, 2010
- Unbiased nonlinear average age-appropriate brain templates from birth to adulthoodNeuroImage, 2009
- Empirical optimization of ASL data analysis using an ASL data processing toolbox: ASLtbxMagnetic Resonance Imaging, 2008
- A theoretical and experimental investigation of the tagging efficiency of pseudocontinuous arterial spin labelingMagnetic Resonance in Medicine, 2007
- Perfusion imagingMagnetic Resonance in Medicine, 1992