Abstract
Twiddler's syndrome, a rare but potentially lethal complication of cardiac pacemaker treatment, is generally diagnosed within the first year of implantation. It is characterized by device malfunction due to dislodgement of cardiac leads resulting from some form of manipulation by the patient. In this report we present a patient who was diagnosed Twiddler's syndrome within the initial 48 h of implantation of permanent pacemaker. In our case, passive fixation of ventricular lead perpetuated this situation and subsequent active fixation prevented any recurrence. Active fixations fixations of device leads are very much essential to prevent this catastrophic complication.Keywords
This publication has 9 references indexed in Scilit:
- Late pacemaker twiddler syndromeClinical Research in Cardiology, 2006
- Twiddler’s Syndrome: An Interesting Cause of Pacemaker FailureCardiology, 2006
- Twiddling to the extreme: development of twiddler syndrome in an implanted cardioverter-defibrilllator.2005
- Early pacemaker twiddler syndromeEP Europace, 2003
- Twiddler's SyndromeNew England Journal of Medicine, 2003
- Defibrillator Twiddler's syndromeThe Annals of Thoracic Surgery, 1995
- Complications of Permanent Transvenous Cardiac Pacing: A 14‐Year Review of all Transvenous Pacemakers Inserted at One Community HospitalPacing and Clinical Electrophysiology, 1987
- [Pacemaker's twiddler syndrome (author's transl)].1979