Variation in CT Number and Image Noise Uniformity According to Patient Positioning in MDCT
- 1 May 2017
- journal article
- research article
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 208 (5) , 1064-1072
Abstract
OBJECTIVE. Many algorithms for clinical decision making rely on assessment of the CT number (expressed as Hounsfield units); however, to our knowledge, few, if any, studies have addressed how CT numbers change as a function of patient positioning within the scanner. MATERIALS AND METHODS. An anthropomorphic phantom underwent imaging with varying amounts of vertical orientation misalignment with respect to isocenter. CT number and noise were measured using ROIs in the upper thorax, mid thorax, and abdomen. The degree of noise nonuniformity and changes in the CT number were assessed by comparing values obtained in the anterior versus posterior ROIs. To add clinical relevance, data on vertical mispositioning were collected from 20,316 clinical abdominal CT scans. Box-and-whisker plot analysis was used to identify the range of patient positioning. RESULTS. Absolute CT number changes of more than 20 HU were observed for some ROIs at phantom positions of 10 cm from isocenter, with important differences noted be...Keywords
This publication has 28 references indexed in Scilit:
- Improvement in CT image resolution due to the use of focal spot deflection and increased samplingJournal of Applied Clinical Medical Physics, 2016
- Hi-Res scan mode in clinical MDCT systems: Experimental assessment of spatial resolution performanceMedical Physics, 2016
- Hepatic Steatosis (Fatty Liver Disease) in Asymptomatic Adults Identified by Unenhanced Low-Dose CTAmerican Journal of Roentgenology, 2010
- Efficient use of automatic exposure control systems in computed tomography requires correct patient positioningActa Radiologica, 2009
- Misoperation of CT Automatic Tube Current Modulation Systems with Inappropriate Patient Centering: Phantom StudiesAmerican Journal of Roentgenology, 2009
- Automatic Patient Centering for MDCT: Effect on Radiation DoseAmerican Journal of Roentgenology, 2007
- Accuracy of the CT numbers of simulated lung nodules imaged with multi-detector CT scannersMedical Physics, 2006
- Adrenal Masses: Characterization with Combined Unenhanced and Delayed Enhanced CTRadiology, 2002
- Characterization of adrenal masses using unenhanced CT: an analysis of the CT literature.American Journal of Roentgenology, 1998
- CT appearance of focal fatty infiltration of the liverAmerican Journal of Roentgenology, 1982