Comparison of 3D-Printed Dental Implants with Threaded Implants for Osseointegration: An Experimental Pilot Study
Open Access
Abstract
This study aimed to compare bone healing and implant stability for three types of dental implants: a threaded implant, a three-dimensional (3D)-printed implant without spikes, and a 3D-printed implant with spikes. In four beagle dogs, left and right mandibular premolars (2nd, 3rd, and 4th) and 1st molars were removed. Twelve weeks later, three types of titanium implants (threaded implant, 3D-printed implant without spikes, and 3D-printed implant with spikes) were randomly inserted into the edentulous ridges of each dog. Implant stability measurements and radiographic recordings were taken every two weeks following implant placement. Twelve weeks after implant surgery, the dogs were sacrificed and bone-to-implant contact (BIC) and bone area fraction occupied (BAFO) were compared between groups. At implant surgery, the primary stability was lower for the 3D-printed implant with spikes (74.05 ± 5.61) than for the threaded implant (83.71 ± 2.90) (p = 0.005). Afterwards, no significant difference in implants’ stability was observed between groups up to post-surgery week 12. Histomorphometrical analysis did not reveal a significant difference between the three implants for BIC (p = 0.101) or BAFO (p = 0.288). Within the limits of this study, 3D-printed implants without spikes and threaded implants showed comparable implant stability measurements, BIC, and BAFO.Keywords
Funding Information
- the Industrial Strategic Technology Development Program - Development of Material Component Technology funded by the Ministry of Trade, Industry & Energy (MOTIE, Korea). (20001221, Development of high strength and fatigue resistance metal and manufacturing technology for root analogue dental implants)
This publication has 28 references indexed in Scilit:
- 3D Printing—Encompassing the Facets of DentistryFrontiers in Bioengineering and Biotechnology, 2018
- 3D printing in dentistryBritish Dental Journal, 2015
- Design and manufacture of customized dental implants by using reverse engineering and selective laser melting technologyThe Journal of Prosthetic Dentistry, 2014
- Current trends in dental implantsJournal of the Korean Association of Oral and Maxillofacial Surgeons, 2014
- Custom-made, root-analogue direct laser metal forming implant: a case reportLasers in Medical Science, 2012
- Immediate Loading of Mandibular Overdentures Supported by Unsplinted Direct Laser Metal‐Forming Implants: Results From a 1‐Year Prospective StudyJournal of Periodontology, 2012
- Direct metal laser sintering (DMLS) of a customized titanium mesh for prosthetically guided bone regeneration of atrophic maxillary archesMedical & Biological Engineering & Computing, 2011
- Direct laser metal sintering as a new approach to fabrication of an isoelastic functionally graded material for manufacture of porous titanium dental implantsDental Materials, 2008
- Immediate Function of Cortically Anchored Disk-Design Implants without Bone Augmentation in Moderately to Severely Resorbed Completely Edentulous MaxillaeJournal of Oral Implantology, 1999
- The Use of Blade Implants in a Selected Population of Partially Edentulous Adults: A Ten‐Year ReportJournal of Periodontology, 1982