Improvement of signal-to-noise and contrast-to-noise ratios in dual-screen computed radiography
Authors:
- Chris C. ShawCSChris C. Shaw
- TienPeng WangTWTienPeng Wang
- Darryl S. BreitensteinDBDarryl S. Breitenstein
- David GurDGDavid Gur
- 1 August 1997
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 24 (8) , 1293-1302
Abstract
A dual-screen computed radiography (CR) technique has been developed to improve and optimize the overall image signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). With this technique, two CR screens are exposed together and separately scanned to form a front and a back image. These two images are then superimposed to form an image of improved SNR and CNR. A mathematical model has been derived to describe the improvement and optimization of the SNR and CNR. Based on this model, the front and back images should be weighted in proportion to their SNR squared to optimize the SNR of the composite image. Imaging experiments have been conducted to verify the theoretical model under mammographic and chest imaging conditions. The results largely agree with the theoretical predictions. It has also been found that optimization of the SNR results in nearly optimal CNR and vice versa. For mammographic imaging, a 14%-22% improvement in the SNR and a 13%-19% improvement in the CNR have been demonstrated. For chest imaging, a 31%-36% improvement in the SNR and a 28%-30% improvement in the CNR has been demonstrated.Keywords
This publication has 14 references indexed in Scilit:
- Physical aspects of photostimulable phosphor computed radiographyBritish Journal of Radiology, 1993
- Portable chest imaging: comparison of storage phosphor digital, asymmetric screen-film, and conventional screen-film systems.Radiology, 1993
- Digital Mammography A Comparison of Two Digitization MethodsInvestigative Radiology, 1989
- Basic imaging properties of a computed radiographic system with photostimulable phosphorsMedical Physics, 1989
- A Comparison of Digitized Storage Phosphors and Conventional Mammography in the Detection of Malignant MicrocalcificationsInvestigative Radiology, 1988
- Imaging performance of a digital storage phosphor systemMedical Physics, 1987
- Excretory urography using computed radiography.Radiology, 1987
- Storage Phosphor System For Computed Radiography: Optical Effects And Detective Quantum Efficiency (DQE).1987
- Storage Phosphor System For Computed Radiography: Screen Optics1986
- Computed radiography utilizing scanning laser stimulated luminescence.Radiology, 1983