Validation of A Portable Beta-Hydroxybutyric Acid Analyser for The Diagnosis of Subclinical Ketosis in Dairy Cows
Authors:
- Kemal AksoyKAKemal Aksoy
- Abdülkerim Deni̇zADAbdülkerim Deni̇z
- Mert Meti̇nMMMert Meti̇n
- Gülten Emek TunaGTGülten Emek Tuna
- 4 September 2023
- journal article
- Published by Kocatepe Veteriner Dergisi in Kocatepe Veterinary Journal
Abstract
On-farm determination of beta-hydroxybutyric acid (BHBA) with an accurate analyser is important for rapid diagnosis of subclinical ketosis (SCK) or clinical ketosis in dairy cows. The accuracy of the Taidoc BHBA portable analyser has not yet been tested and validated. To validate the Taidoc BHBA analyser with another device Wellion (already validated) in 84 dairy cows. Whole blood was collected from the coccygeal vein and immediately analysed with Taidoc and Wellion in one drop of whole venous blood of 84 dairy cows. The averages of BHBA concentration were 0.79±0.66 and 1.09±0.51 mmol/L for Wellion and Taidoc, respectively. The correlation coefficient was r=0.64 (pKeywords