Performance of an ICD algorithm to detect lead noise and reduce inappropriate shocks
- 6 January 2016
- journal article
- Published by Springer Nature in Journal of Interventional Cardiac Electrophysiology
- Vol. 45 (2) , 225-232
Abstract
Implantable cardioverter-defibrillators (ICD) provide treatment for life-threatening ventricular tachyarrhythmias. Failure of the pace/sense conductor of an ICD lead can cause noise on the sensing electrogram (EGM) that may be misinterpreted as ventricular activity, triggering inappropriate therapy. An algorithm based upon the confirmation of ventricular activity from a far-field EGM has been developed to reduce inappropriate therapies resulting from this type of lead failure, while ensuring that appropriate therapy is delivered. The objectives of this study were to evaluate the algorithm’s ability to discriminate lead noise from ventricular tachycardia/ventricular fibrillation (VT/VF) and to determine whether it inhibits inappropriate shocks without delaying appropriate shocks. The algorithm was prospectively tested using near- and far-field EGM recordings from patients in three conditions: normal sinus rhythm with sustained and non-sustained lead noise via manipulation of the ICD pocket or lead system, and VT/VF induced during defibrillation threshold testing. The recordings were played through a bench-top device running the algorithm with the diagnosis, time to diagnosis, and inhibition of therapy documented. The algorithm detected noise and withheld inappropriate therapy in 231 of 238 recordings of sustained lead noise that would otherwise have been diagnosed as VT/VF (97.1 %). Non-sustained lead noise was correctly diagnosed in 47 of the 52 recordings (90.4 %). The device appropriately identified all 853 recordings of VT/VF (100 %), without an increase in the time to detection (0.01 ± 0.14 s). The SecureSenseTM algorithm correctly diagnosed sustained and non-sustained lead noise recordings without compromising detection of VT/VF. Use of the algorithm may reduce inappropriate shocks and alert clinicians to lead noise indicative of lead failure.Keywords
This publication has 22 references indexed in Scilit:
- Single Channel ICD Noise Without Loss of Conductor Circuit IntegrityPacing and Clinical Electrophysiology, 2011
- Downloadable Algorithm to Reduce Inappropriate Shocks Caused by Fractures of Implantable Cardioverter-Defibrillator LeadsCirculation, 2008
- Prognostic Importance of Defibrillator Shocks in Patients with Heart FailureNew England Journal of Medicine, 2008
- Inappropriate Implantable Cardioverter-Defibrillator Shocks in MADIT IIJACC, 2008
- Development and testing of an algorithm to detect implantable cardioverter-defibrillator lead failureHeart Rhythm, 2006
- Amiodarone or an Implantable Cardioverter–Defibrillator for Congestive Heart FailureThe New England journal of medicine, 2005
- Prophylactic Implantation of a Defibrillator in Patients with Myocardial Infarction and Reduced Ejection FractionThe New England journal of medicine, 2002
- Randomized Comparison of Antiarrhythmic Drug Therapy With Implantable Defibrillators in Patients Resuscitated From Cardiac ArrestCirculation, 2000
- Canadian Implantable Defibrillator Study (CIDS)Circulation, 2000
- A Comparison of Antiarrhythmic-Drug Therapy with Implantable Defibrillators in Patients Resuscitated from Near-Fatal Ventricular ArrhythmiasNew England Journal of Medicine, 1997