Accuracy of capillary blood glucose for 50-g glucose challenge test for gestational diabetes screening
- 7 February 2022
- journal article
- research article
- Published by Springer Nature in Diabetology International
- Vol. 13 (3) , 561-565
Abstract
Purpose To assess the accuracy of capillary blood glucose (CBG) compared to conventional venous plasma glucose (VPG) testing for 50-g glucose challenge test (GCT) in gestational diabetes (GDM) screening. Methods A total of 300 women were enrolled and 50-g GCT for GDM screening was offered. At 1 h after glucose loading, CBG was evaluated by CONTOUR® PLUS glucose meter by well-trained nurses immediately after venipuncture for VPG. Results of CBG were compared with those from VPG to evaluate its accuracy. Women with venous plasma glucose > 140 mg/dL were offered 100-g OGTT for GDM diagnosis. Results The mean age was 30.2 years and the mean gestational age at testing was 21.8 weeks. GDM was diagnosed in 34 women (11.3%). The mean VPG was 142.1 ± 32.9 mg/dL and the mean CBG was 129.3 ± 33.5 mg/dL. Mean difference was −12.3 ± 12.5 mg/dL, corresponding to −8.8 ± 11.4%. CBG significantly correlated with VPG with correlation coefficient of 0.929, p < 0.001. In the detection of abnormal 50-g GCT results (VPG ≥ 140 mg/dL), at 126 mg/dL cutoff, CBG had sensitivity of 92.5%, specificity of 81.8%, and positive and negative predictive values of 82.8%and 92%. None of the GDM would have been missed if CBG was used. Conclusion CBG by a certified glucose meter could be considered as an alternative to conventional VPG testing for 50-g GCT for GDM screening using 126 mg/dL cutoff value.Keywords
This publication has 16 references indexed in Scilit:
- Prevalence of gestational diabetes mellitus based on various screening strategies in western Kenya: a prospective comparison of point of care diagnostic methodsBMC Pregnancy and Childbirth, 2017
- Evaluation of point‐of‐care maternal glucose measurements for the diagnosis of gestational diabetes mellitusBJOG: An International Journal of Obstetrics & Gynaecology, 2016
- Accuracy and precision assessment of a new blood glucose monitoring systemClinical Chemistry and Laboratory Medicine (CCLM), 2016
- Accuracy Evaluation of CONTOUR®PLUS Compared With Four Blood Glucose Monitoring SystemsDiabetes Therapy, 2015
- Use of capillary blood glucose for screening for gestational diabetes mellitus in resource-constrained settingsActa Diabetologica, 2015
- Accuracy and User Performance Evaluation of a Blood Glucose Monitoring SystemDiabetes Technology & Therapeutics, 2015
- Comparison of Venous Plasma Glucose and Capillary Whole Blood Glucose in the Diagnosis of Gestational Diabetes Mellitus: A Community-Based StudyDiabetes Technology & Therapeutics, 2012
- Performance of the Roche Accu-Chek Active Glucose Meter to Screen for Gestational Diabetes Mellitus Using Fasting Capillary BloodDiabetes Technology & Therapeutics, 2011
- Capillary blood glucose screening (Accu-Chek Advantage) for gestational diabetes.2009
- Comparison of point-of-care-testing glucose meters with standard laboratory measurement of the 50g-glucose-challenge test (GCT) during pregnancyClinical Biochemistry, 2003