Off‐axis dose response characteristics of an amorphous silicon electronic portal imaging device
- 17 September 2007
- journal article
- Published by Wiley in Medical Physics
- Vol. 34 (10) , 3815-3824
Abstract
Amorphous silicon electronic portal imaging devices (EPIDs) have typically been calibrated to dose at central axis (CAX). Division of acquired images by the flood‐field (FF) image that corrects for pixel sensitivity variation as well as open field energy‐dependent off‐axis response variation should result in a flat EPID response over the entire matrix for the same field size. While the beam profile can be reintroduced to the image by an additional correction matrix, the CAX EPID response to dose calibration factor is assumed to apply to all pixels in the detector. The aim of this work was to investigate the dose response of the Varian aS500 amorphous silicon detector across the entire detector area. First it was established that the EPID response across the panel became stable (within ) for MU settings greater than . The EPID was then FF calibrated with a high MU setting of for all subsequent experiments. Whole detector images with varying MU settings from 2–500 were then acquired for two dose rates (300 and ) for photons for two EPIDs. The FF corrected EPID response was approximately flat or uniform across the detector for greater than delivered (within 0.5%). However, the off‐axis EPID response was greater than the CAX response for small MU irradiations, giving a raised EPID profile. Up to 5% increase in response at off‐axis compared to CAX was found for very small MU settings for one EPID, while it was within 2% for the second (newer) EPID. Off‐axis response nonuniformities attributed to detector damage were also found for the older EPID. Similar results were obtained with the EPID at energy and operating in asynchronous mode (acquisition not synchronized with beam pulses), however the profiles were flatter and more irregular for the small MU irradiations. By moving the detector laterally and repeating the experiments, the increase in response off‐axis was found to depend on the pixel position relative to the beam CAX. When the beam was heavily filtered by a phantom the off‐axis response variation was reduced markedly to within 0.5% for all MU settings. Independent measurements of off‐axis point doses with ion chamber did not show any change in off‐axis factor with MUs. Measurements of beam quality for MU settings of 2, 5, and 100 at central axis and at off‐axis could not explain the effect. The response change is unlikely to be significant for clinical IMRT verification with this imaging/acclerator system where MUs are of the order of 100–300, provided the detector does not exhibit radiation damage artifacts.Keywords
This publication has 15 references indexed in Scilit:
- Monte Carlo modelling of a‐Si EPID response: The effect of spectral variations with field size and positionMedical Physics, 2006
- Comparison of ghosting effects for three commercial EPIDsMedical Physics, 2006
- Correction of pixel sensitivity variation and off‐axis response for amorphous silicon EPID dosimetryMedical Physics, 2005
- Improving IMRT quality control efficiency using an amorphous silicon electronic portal imagerMedical Physics, 2005
- Dose‐response characteristics of an amorphous silicon EPIDMedical Physics, 2005
- Consequences of the spectral response of an a‐Si EPID and implications for dosimetric calibrationMedical Physics, 2005
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapyMedical Physics, 2003
- Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging deviceMedical Physics, 2001
- Relative dosimetry using active matrix flat-panel imager (AMFPI) technologyMedical Physics, 1999
- X-ray quantum limited portal imaging using amorphous silicon flat-panel arraysMedical Physics, 1998