Potential of combining iterative reconstruction with noise efficient detector design: aggressive dose reduction in head CT
- 1 June 2015
- journal article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 88 (1050) , 20140404
Abstract
With further increase of CT numbers and their dominant contribution to medical exposure, there is a recent quest for more effective dose control. While reintroduction of iterative reconstruction (IR) has proved its potential in many applications, a novel focus is placed on more noise efficient detectors. Our purpose was to assess the potential of IR in combination with an integrated circuit detector (ICD) for aggressive dose reduction in head CT. Non-contrast low-dose head CT [190 mAs; weighted volume CT dose index (CTDIvol), 33.2 mGy] was performed in 50 consecutive patients, using a new noise efficient detector and IR. Images were assessed in terms of quantitative and qualitative image quality and compared with standard dose acquisitions (320 mAs; CTDIvol, 59.7 mGy) using a conventional detector and filtered back projection. By combining ICD and IR in low-dose examinations, the signal to noise was improved by about 13% above the baseline level in the standard-dose control group. Both, contrast-to-noise ratio (2.02 ± 0.6 vs 1.88 ± 0.4; p = 0.18) and objective measurements of image sharpness (695 ± 84 vs 705 ± 151 change in Hounsfield units per pixel; p = 0.79) were fully preserved in the low-dose group. Likewise, there was no significant difference in the grading of several subjective image quality parameters when both noise-reducing strategies were used in low-dose examinations. Combination of noise efficient detector with IR allows for meaningful dose reduction in head CT without compromise of standard image quality. Our study demonstrates the feasibility of almost 50% dose reduction in head CT dose (1.1 mSv per scan) through combination of novel dose-reducing strategies.Keywords
This publication has 27 references indexed in Scilit:
- Added Value of Integrated Circuit Detector in Head CTAcademic Radiology, 2014
- Integrated Circuit Detector Technology in Abdominal CT: Added Value in Obese PatientsAmerican Journal of Roentgenology, 2014
- Adaptive statistical iterative reconstruction reduces patient radiation dose in neuroradiology CT studiesNeuroradiology, 2014
- Reducing Image Noise in Computed Tomography (CT) ColonographyJournal of Computer Assisted Tomography, 2014
- Iterative Reconstruction to Preserve Image Quality and Diagnostic Accuracy at Reduced Radiation Dose in Coronary CT AngiographyJACC: Cardiovascular Imaging, 2013
- Electronic Noise in CT Detectors: Impact on Image Noise and ArtifactsAmerican Journal of Roentgenology, 2013
- How far can the radiation dose be lowered in head CT with iterative reconstruction? Analysis of imaging quality and diagnostic accuracyEuropean Radiology, 2013
- Stenosis Quantification in Coronary CT AngiographyInvestigative Radiology, 2013
- Effect of low doses of ionising radiation in infancy on cognitive function in adulthood: Swedish population based cohort studyBMJ, 2004
- Managing patient dose in computed tomographyAnnals of the ICRP, 2000