PVDF-Based Piezoelectric Microphone for Sound Detection Inside the Cochlea: Toward Totally Implantable Cochlear Implants
Open Access
- 1 January 2018
- journal article
- research article
- Published by SAGE Publications in Trends in Hearing
- Vol. 22
Abstract
We report the fabrication and characterization of a prototype polyvinylidene fluoride polymer-based implantable microphone for detecting sound inside gerbil and human cochleae. With the current configuration and amplification, the signal-to-noise ratios were sufficiently high for normally occurring sound pressures and frequencies (ear canal pressures >50–60 dB SPL and 0.1–10 kHz), though 10 to 20 dB poorer than for some hearing aid microphones. These results demonstrate the feasibility of the prototype devices as implantable microphones for the development of totally implantable cochlear implants. For patients, this will improve sound reception by utilizing the outer ear and will improve the use of cochlear implants.Keywords
This publication has 31 references indexed in Scilit:
- High performance piezoelectric devices based on aligned arrays of nanofibers of poly(vinylidenefluoride-co-trifluoroethylene)Nature Communications, 2013
- A systematic review of the effectiveness and cost‐effectiveness of bilateral multichannel cochlear implants in adults with severe‐to‐profound hearing lossClinical Otolaryngology, 2012
- Porous PVDF As Effective Sonic Wave Driven NanogeneratorsNano Letters, 2011
- Piezoelectric materials mimic the function of the cochlear sensory epitheliumProceedings of the National Academy of Sciences, 2011
- Microphone based on Polyvinylidene Fluoride (PVDF) micro-pillars and patterned electrodesSensors and Actuators A: Physical, 2009
- Differential Intracochlear Sound Pressure Measurements in Normal Human Temporal BonesJournal of the Association for Research in Otolaryngology, 2008
- Cochlear implants: Current designs and future possibilitiesJournal of Rehabilitation Research and Development, 2008
- Cochlear implants: A remarkable past and a brilliant futureHearing Research, 2008
- A novel wearable sensor device with conductive fabric and PVDF film for monitoring cardiorespiratory signalsSensors and Actuators A: Physical, 2006
- The possibility of formation of large ferroelectric domains in a copolymer of vinylidene fluoride and trifluoroethyleneJournal of Applied Physics, 1992