Abstract
Corrosion of modular metal‐on‐metal acetabular tapers in total hip arthroplasty (THA) systems is often attributed to mechanically driven processes. Recent findings suggest that mechanically assisted crevice corrosion (MACC) might not be the dominant cause of corrosion in shell‐liner tapers. This study aims to document and present the corrosion modes observed in metal–metal acetabular liners. Twenty‐one retrieved wrought CoCrMo liners were examined using digital optical microscopy (DOM), scanning electron microscopy (SEM) and energy dispersive x‐ray spectroscopy (EDS). Corrosion‐related damage was documented in nonengagement taper regions, outside of direct taper contact. Within engagement regions, nonmechanically driven corrosion features (pitting, intergranular corrosion) were observed adjacent to fretting and material transfer, which rely on mechanical contact; corrosion independent of MACC was observed even in contact regions. Corrosion types observed included intergranular corrosion (IGC), pitting attack, phase boundary dissolution, all both outside and inside of taper junctions, and MACC within contact regions of the taper. Typical fretting scars associated with MACC were mostly absent, and were not always associated with corrosion damage where present. Finally, hard phase particles (Mo–Si–O) released from the wrought CoCrMo microstructure had redeposited within regions with material loss. Acetabular taper corrosion modes differ significantly from those in head–neck tapers and are dominated by electrochemically driven processes, not mechanical processes, as indicated by corrosion in noncontact regions. With greater prevalence of dual mobility hip implants, acetabular taper corrosion processes must be understood in order to limit their impact on device performance.Keywords
This publication has 45 references indexed in Scilit:
- Is Increased Modularity Associated With Increased Fretting and Corrosion Damage in Metal-On-Metal Total Hip Arthroplasty Devices?The Journal of Arthroplasty, 2013
- Poor Short Term Outcome with a Metal-On-Metal Total Hip ArthroplastyThe Journal of Arthroplasty, 2013
- Retrieval Analysis of Failed Constrained Acetabular LinersThe Journal of Arthroplasty, 2009
- Soluble and particulate Co‐Cr‐Mo alloy implant metals activate the inflammasome danger signaling pathway in human macrophages: A novel mechanism for implant debris reactivityJournal of Orthopaedic Research, 2008
- The effect of nano- and micron-sized particles of cobalt–chromium alloy on human fibroblasts in vitroBiomaterials, 2007
- Characterization of HLA-A2-restricted HPV-16 E7-specific CD8+ T-cell immune responses induced by DNA vaccines in HLA-A2 transgenic miceGene Therapy, 2005
- ROS to the rescueAmerican Journal of Physiology-Lung Cellular and Molecular Physiology, 2004
- Changes in the surface oxide composition of Co–Cr–Mo implant alloy by macrophage cells and their released reactive chemical speciesBiomaterials, 2004
- Fretting corrosion accelerates crevice corrosion of modular hip tapersJournal of Applied Biomaterials, 1995
- In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy couplingJournal of Biomedical Materials Research, 1993