Abstract
We write to inform your readers about our experience with the configuration of the Dräger Apollo anesthesia machine (Dräger Medical, Inc., Telford, PA). We have had several instances where end-tidal carbon dioxide readings were skewed or undetected by the gas analyzer after successful tracheal intubation. Upon inspection of the source of the problem, it was found that the water trap (positioned just below the isoflurane vaporizer; fig. 1) was severely cracked and disfigured (fig. 2). This compromise of the water trap led to the inability to accurately detect and monitor end-tidal carbon dioxide. Upon replacement of the water trap, accurate carbon dioxide monitoring was once again possible.