Development of an in-vacuum long trace profiler for in situ measurement of optical components
- 1 September 2025
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 96 (9)
Abstract
In this work, we developed a vacuum-compatible long trace profiler (LTP) for in situ metrology of ultra-precise x-ray optics within synchrotron vacuum chambers. Although traditional LTPs operate ex situ under atmospheric pressure, earlier optical setups—such as that by Qian et al.—performed in situ distortion measurements by directing laser beams through vacuum viewports. While these configurations enabled in situ monitoring of mirror deformation, their accuracy was constrained by optical distortions from vacuum windows and by beam deviations caused by air turbulence. To overcome these limitations, we developed a fully in-vacuum LTP system installed directly inside the vacuum chamber, fully compatible with ultrahigh vacuum (UHV) conditions (<1 × 10−8 Torr). Based on the original Takacs LTP design, the system incorporates a vacuum-compatible CMOS sensor, an external laser to reduce heat and weight, and a motorized four-axis alignment stage. This system is designed to measure low-spatial-frequency slope error, which is critical for evaluating optical figure quality. Performance was validated under identical atmospheric conditions using a conventional LTP for comparison. Under identical conditions, the in-vacuum LTP measured slope errors of 100 nrad RMS (flat) and 200 nrad RMS (curved). However, measurable deviations in curvature radius and coma coefficient were observed, attributed to stage wobble and lens aberrations. These findings confirm baseline performance and demonstrate the system’s feasibility for UHV-compatible slope metrology. The in-vacuum LTP is intended to support real-time slope monitoring and feedback correction of active optical components during operation at synchrotron beamlines.Keywords
Funding Information
- National Synchrotron Radiation Research Center (11362PT35)
This publication has 15 references indexed in Scilit:
- Towards a new generation long trace profiler LTP-2020: optical design of pencil beam interferometry sensor2023
- Application of a long-trace profiler in situ at TPS beamlinesJournal of Physics: Conference Series, 2022
- Development of the Soft X-ray AGM–AGS RIXS beamline at the Taiwan Photon SourceJournal of Synchrotron Radiation, 2021
- Raytracing the long trace profiler2020
- Development of a long trace profiler
in situ
at National Synchrotron Radiation Research CenterReview of Scientific Instruments, 2019
- A submicron soft x-ray active grating monochromator beamline for ultra-high resolution angle-resolved photoemission spectroscopyAIP Conference Proceedings, 2019
- A high precision 25-actuator optical surface bender for ultra-high resolution soft x-ray spectroscopiesAIP Conference Proceedings, 2019
- Wave front-splitting phase shift beam splitter for pencil beam interferometerReview of Scientific Instruments, 2003
- Portable long trace profiler: Concept and solutionReview of Scientific Instruments, 2001
- Advantages of in-situ LTP distortion profile test on high heat load mirrors and applications1996