Abstract
Fused‐Core particles have recently been introduced as an alternative to using sub‐2‐μm particles in chromatographic separations. Fused‐Core particles are composed of a 1.7 μm solid core surrounded by a 0.5 μm porous silica layer (dp = 2.7 μm) to reduce mass transfer and increase peak efficiency. The performance of two commercially available Fused‐Core particles (Advanced Materials Technology Halo C18 and Supelco Ascentis Express C18) was compared with sub‐2‐μm particles from Waters, Agilent, and Thermo Scientific. Although the peak efficiencies were only ∼︁80% of those obtained by the Waters Acquity particles, the 50% lower backpressure allowed columns to be coupled in series to increase peak efficiency to 92 750 plates. The low backpressure and high efficiencies of the Fused‐Core particles offer a viable alternative to using sub‐2‐μm particles and very‐high‐pressure LC instrumentation.

This publication has 13 references indexed in Scilit: