Using the Allan variance and power spectral density to characterize DC nanovoltmeters
Author:
- 1 April 2001
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 50 (2) , 445-448
Abstract
When analyzing nanovoltmeter measurements, stochastic serial correlations are often ignored and the experimental standard deviation of the mean is assumed to be the experimental standard deviation of a single observation divided by the square root of the number of observations. This is justified only for white noise. This paper demonstrates the use of the power spectrum and the Allan variance to analyze data, identify the regimes of white noise, and characterize the performance of digital and analog DC nanovoltmeters. Limits imposed by temperature variations, 1/f noise and source resistance are investigated.Keywords
This publication has 6 references indexed in Scilit:
- Using power spectra and Allan variancesto characterisethe noise of Zener-diode voltage standardsIEE Proceedings - Science, Measurement and Technology, 2000
- TESTING FOR CORRELATIONS IN MEASUREMENTS2000
- Measurements of the temperature dependenceof the output voltages of someZener diode based voltage standardsIEE Proceedings - Science, Measurement and Technology, 1998
- Low-frequency spectral analysis of DC nanovoltmeters and voltage reference standardsIEEE Transactions on Instrumentation and Measurement, 1997
- Characterization of frequency stability in precision frequency sourcesProceedings of the IEEE, 1991
- Should the classical variance be used as a basic measure in standards metrology?IEEE Transactions on Instrumentation and Measurement, 1987