A programmable and self-adjusting class E amplifier for efficient wireless powering of biomedical implants
Authors:
- 1 August 2015
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE) in 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
- Vol. 2015 (1094687X) , 3193-3196
Abstract
In this paper, an enhanced approach of a class E amplifier being insensitive to coil impedance variations is presented. While state of the art class E amplifiers widely being used to supply implanted systems show a strong degradation of efficiency when powering distance, coil orientation or the implant current consumption deviate from the nominal design, the presented concept is able to detect these deviations on-line and to reconfigure the amplifier automatically. The concept is facilitated by a new approach of sensing the load impedance without interruption of the power supply to the implant, while the main components of the class E amplifier are programmable by software. Therefore, the device is able to perform dynamic impedance matching. Besides presenting the operational principle and the design equations, we show an adaptive prototype reader system which achieves a drain efficiency of up to 92% for a wide range of reflected coil impedances from 1 to 40 Ω. The integrated communication concept allows downlink data rates of up to 500 kBit/s, while the load modulation based uplink from implant to reader was verified of providing up to 1.35 MBit/s.Keywords
This publication has 9 references indexed in Scilit:
- Efficient inductive powering of brain implanted sensors2015
- Design of Single-Switch Inverters for Variable Resistance/Load Modulation OperationIEEE Transactions on Power Electronics, 2014
- Tuning Class E Inverters Applied in Inductive Links Using Saturable ReactorsIEEE Transactions on Power Electronics, 2013
- A 232-channel retinal vision prosthesis with a miniaturized hermetic package2012
- Inductive power link for a wireless cortical implant with biocompatible packaging2010
- The concepts of inductive powering2009
- Feedback Analysis and Design of RF Power Links for Low-Power Bionic SystemsIEEE Transactions on Biomedical Circuits and Systems, 2007
- A Low-Power Integrated Circuit for a Wireless 100-Electrode Neural Recording SystemIEEE Journal of Solid-State Circuits, 2006
- Idealized operation of the class E tuned power amplifierIEEE Transactions on Circuits and Systems, 1977