Mechanical and Corrosion Testing of Magnesium WE43 Specimens for Pitting Corrosion Model Calibration
- 16 July 2018
- journal article
- research article
- Published by Wiley in Advanced Engineering Materials
- Vol. 20 (10)
Abstract
Computational modeling can play an important role in the analysis, design, and development of complex medical devices such as biodegradable coronary stents. In this study, experimental mechanical and corrosion testing is conducted to characterize the mechanical and corrosion behavior of magnesium WE43 alloy, a candidate base material for biodegradable magnesium alloy stents. Previously developed uniform and pitting corrosion models are calibrated based on in vitro mechanical and corrosion testing of magnesium WE43 alloy specimens. The calibrated pitting corrosion model can capture the mechanical and corrosion behavior of magnesium WE43, including the experimentally observed non‐linear reduction in failure strength with mass loss, whereas the uniform corrosion model is incapable of capturing this trend. The calibrated corrosion models will be used in future research on magnesium alloy stents.Keywords
Funding Information
- Irish Research Council for Science, Engineering and Technology
This publication has 36 references indexed in Scilit:
- Bioresorbable Drug-Eluting Magnesium-Alloy Scaffold for Treatment of Coronary Artery DiseaseInternational Journal of Molecular Sciences, 2013
- Tensile testing as a novel method for quantitatively evaluating bioabsorbable material degradationJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2012
- A corrosion model for bioabsorbable metallic stentsActa Biomaterialia, 2011
- Research on an Mg–Zn alloy as a degradable biomaterialActa Biomaterialia, 2010
- Influence of grain size on mechanical and corrosion properties of magnesium alloy for medical implantsMaterialwissenschaft und Werkstofftechnik, 2009
- Early- and Long-Term Intravascular Ultrasound and Angiographic Findings After Bioabsorbable Magnesium Stent Implantation in Human Coronary ArteriesJACC: Cardiovascular Interventions, 2009
- In vitro corrosion and biocompatibility of binary magnesium alloysBiomaterials, 2009
- A Possible Biodegradable Magnesium Implant MaterialAdvanced Engineering Materials, 2007
- Understanding Magnesium Corrosion—A Framework for Improved Alloy PerformanceAdvanced Engineering Materials, 2003
- Fracture behavior of AZ91 magnesium alloyMaterials Letters, 2000