Short Communication: The 2005 Italian Quality Control Study for the Evaluation of CD4 Cells in Centers Involved in the Treatment of HIV Type 1 Patients
Authors:
- Federico MartiniFMFederico Martini
- Enrico GirardiEGEnrico Girardi
- Eleonora CiminiECEleonora Cimini
- The Other Members Of The 2005 Italian Cd4 Quality Control NetworkTNThe Other Members Of The 2005 Italian Cd4 Quality Control Network
- 1 June 2007
- journal article
- research article
- Published by SAGE Publications in AIDS Research and Human Retroviruses
- Vol. 23 (6) , 777-781
Abstract
We report the results of an external quality control program, including 17 Italian centers involved in the care of patients infected by HIV, to evaluate CD4 T cell count proficiency and reproducibility. The centers received two commercial stabilized blood preparations, one with “normal” and one with “low” CD4 T cell content. The centers were asked to process the samples two times, 1 week apart, with the same procedure used for samples from HIV patients. Most centers showed a good performance of CD4 frequency and absolute count determinations. In particular, the “low” sample was correctly analyzed by all centers; only two underestimated the “normal” sample CD4 frequency, and only one underestimated the CD4 absolute count by >100 CD4 cells/μl. Overall, our data suggest that most Italian laboratories provide reliable and reproducible results in evaluating CD4 T cells in HIV+ samples.Keywords
This publication has 16 references indexed in Scilit:
- Impact of the international program for quality assessment and standardization for immunological measures relevant to HIV/AIDS: QASICytometry, 2002
- Impact of unified procedures as implemented in the Canadian Quality Assurance Program for T lymphocyte subset enumerationCytometry, 1998
- Analysis of variation in results of flow cytometric lymphocyte immunophenotyping in a multicenter studyCytometry, 1997
- Assessment of Interlaboratory Variability of ImmunophenotypingAnnals of the New York Academy of Sciences, 1993
- NIAID Division of AIDS Flow Cytometry Quality Assessment ProgramAnnals of the New York Academy of Sciences, 1993
- Development of Clinical Standards for Flow CytometryAnnals of the New York Academy of Sciences, 1993
- Assessment of the Effects of Instrumentation, Monoclonal Antibody, and Fluorochrome on Flow Cytometric Immunophenotyping: A Report Based on 2 Years of the NIAID DAIDS Flow Cytometry Quality Assessment ProgramClinical Immunology and Immunopathology, 1993
- Nationwide quality control trial on lymphocyte immunophenotyping and flow cytometer performance in ItalyCytometry, 1993
- Quality control in the flow cytometric measurement of T-lymphocyte subsets: The Multicenter AIDS Cohort Study experienceClinical Immunology and Immunopathology, 1990
- The Prognostic Value of Cellular and Serologic Markers in Infection with Human Immunodeficiency Virus Type 1New England Journal of Medicine, 1990