A Study on Measurement Uncertainty of a Vortex Flow Meter in Discrete Liquid Phase
Authors:
- 1 September 2011
- journal article
- Published by Trans Tech Publications, Ltd. in Advanced Materials Research
- Vol. 346, 593-599
Abstract
The measurement uncertainty of vortex flowmeter was examined when the gas flow measured was injected with liquid. This test was performed in a multiphase flow calibration facility. A vortex flowmeter of 50 mm in diameter was installed in a 100 mm test section. The gas volume flow rate was held in 141m3/h and the liquid was injected into the gas flow. The liquid volume fractions used at the gas volume flow rate were 0.0106%, 0.0213%,0.0355%,0.0496%,0.0638%,0.0780% and 0.0922%. The small amount of liquid in the gas as discrete droplet is called discrete liquid phase. Analysis on the vortex shedding frequency obtained from a frequency spectra showed that the strouhal values changed from 0.305 to 0.385 with the discrete liquid phase increasing and the total uncertainty of vortex flowmeter was from 0.869% to 2.196%. The experimental result can supply experimental basis for the measurement error correction of vortex flowmeter worked in gas flow with discrete liquid phase.Keywords
This publication has 5 references indexed in Scilit:
- On measurement uncertainty of a vortex flowmeterFlow Measurement and Instrumentation, 2005
- The vortex flowmeter: various methods of investigating phenomenaMeasurement Science and Technology, 2005
- Flow measurement by a new type vortex flowmeter of dual triangulate bluff bodySensors and Actuators A: Physical, 2004
- Measurement uncertainty according to ISO/BIPM-GUMThermochimica Acta, 2002
- On the application of the guide to the expression of uncertainty in measurement to measuring instrumentsMeasurement, 1995