A Beating Heart Testbed for the Evaluation of Robotic Cardiovascular Interventions
Authors:
- 1 August 2018
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1076-1082
Abstract
The improved natural hemodynamics offered by mitral valve (MV) repair strategies aims to prevent heart failure and to minimize the use of long-term anticoagulant. This combined with the reduced patient trauma offered by minimally invasive surgical (MIS) interventions, requires an increase in capabilities of MIS MV repair. The use of robotic catheters have been described in MIS applications such as navigational tasks, ablation and MV repair. The majority of the robotic catheters are evaluated in testbeds capable of partially mimicking the cardiac environment (e.g., beating heart motion or relevant anatomy), while the validation of robotic catheters in a clinical scenario is associated with significant preparation time and limited availability. Therefore, continuous catheter development could be aided by an accessible and available testbed capable of reproducing beating heart motions, circulation and the relevant anatomy in MIS cardiovascular interventions. In this study, we contribute a beating heart testbed for the evaluation of robotic catheters in MIS cardiovascular interventions. Our work describes a heart model with relevant interior structures and an integrated realistic MV model, which is attached to a Stewart platform in order to reproduce the beating heart motions based on pre-operative patient data. The beating heart model is extended with an artificial aortic valve, a systemic arterial model, a venous reservoir and a pulsatile pump to mimic the systemic circulation. Experimental evaluation showed systemic circulation and beating heart motion reproduction for 70 BPM with a mean absolute distance error of 1.26 mm, while a robotic catheter in the heart model is observed by ultrasound imaging and electromagnetic position tracking. Therefore, the presented testbed is capable of evaluating MIS robotic cardiovascular interventions such as MV repair, navigation tasks and ablation.Keywords
This publication has 25 references indexed in Scilit:
- Robotic catheter cardiac ablation combining ultrasound guidance and force controlThe International Journal of Robotics Research, 2014
- A Novel Left Heart Simulator for the Multi-modality Characterization of Native Mitral Valve Geometry and Fluid MechanicsAnnals of Biomedical Engineering, 2012
- Percutaneous intracardiac beating-heart surgery using metal MEMS tissue approximation toolsThe International Journal of Robotics Research, 2012
- Predictors of Mitral Valve Repair: Clinical and Surgeon FactorsThe Annals of Thoracic Surgery, 2010
- Development of a Physical Windkessel Module to Re-Create In Vivo Vascular Flow Impedance for In Vitro ExperimentsCardiovascular Engineering and Technology, 2010
- A Novel Approach to the In Vitro Hydrodynamic Study of the Aortic Valve: Mock Loop Development and TestASAIO Journal, 2010
- Robot-assisted Active Catheter Insertion: Algorithms and ExperimentsThe International Journal of Robotics Research, 2009
- Robot-assisted catheter manipulation for intracardiac navigationInternational Journal of Computer Assisted Radiology and Surgery, 2009
- The arterial WindkesselMedical & Biological Engineering & Computing, 2008
- Minimally Invasive, Robotic, and Off-Pump Mitral Valve SurgerySeminars in Thoracic and Cardiovascular Surgery, 2006