Linear optical coherence tomography system with a downconverted fringe pattern
- 15 July 2004
- journal article
- research article
- Published by Optica Publishing Group in Optics Letters
- Vol. 29 (14) , 1644-1646
Abstract
Linear optical coherence tomography (LOCT) systems are a simple and robust alternative to time-domain optical coherence tomography systems, but a detector with approximately pixels is needed for an imaging depth of 2 mm. We present a new system for LOCT with a special mask attached to the image sensor. The mask essentially performs a downconversion of the spatial frequencies by multiplication with a second spatial frequency. This reduces the fringe frequency of the optical coherence tomography signal so that the signal can be sampled with fewer pixels.
Keywords