Abstract
We introduce a method to measure radio frequency (RF) electric fields (E-fields) using atoms contained in a prism-shaped vapor cell. The method utilizes the concept of electromagnetically induced transparency with Rydberg atoms. The RF E-field induces changes in the index of refraction of the vapor resulting in deflection of the probe laser beam as it passes through the prism-shaped vapor cell. We measured a minimum RF E-field of 8.25 with a sensitivity of . The experimental results agree with a numerical model that includes dephasing effects. We discuss possible improvements to obtain higher sensitivity for RF E-field measurements.
Funding Information
  • Defense Advanced Research Projects Agency (60181-PH-DRP)
  • Carl-Zeiss-Stiftung

This publication has 45 references indexed in Scilit: