Comparison between two hand-held autorefractors: the Sure-Sight and the Retinomax
- 1 January 2004
- journal article
- research article
- Published by Taylor & Francis in Strabismus
- Vol. 12 (4) , 261-274
Abstract
Purpose To compare the results of manifest refraction obtained with two different hand-held autorefractors (Sure-Sight, Welch Allyn Co. and Retinomax 1, Nikon Inc.) and with the Topcon RMA 6000 on-table autorefractor in order to estimate any potential bias between these refractometers and to compare the diagnostic performances of these two hand-held autorefractors as screening devices. methods Ninety-eight children were refracted under manifest conditions with the three above-mentioned refractometers and under cycloplegic conditions with the Topcon on-table autorefractor, or by means of retinoscopy. The agreement between the manifest measurements obtained with the three different autorefractors was studied using the method of Bland and Altman. The validity of several thresholds of manifest refractive anomalies as measured with the Sure-Sight and with the Retinomax was estimated by receiver operating characteristic (ROC) curves using cycloplegic measures as reference. results There is a spherical positive bias of 1 D between the Sure-Sight and the Retinomax and better agreement between the Topcon and the Retinomax. The surface area indexes of the ROC curves and the diagnostic performances in term of sensitivity and specificity are better with the Retinomax in cases of hyperopia, astigmatism and anisometropia. For myopia, the Sure-Sight has better performance. conclusions The results suggest that either device may be sufficient for assessing refractive errors in children in a screening setting. However, because of a bias between both refractometers in measuring the sphere, distinctive referral criteria must be chosen for the detection of hyperopia and myopia. The diagnostic performance is slightly in favor of the Retinomax.Keywords
This publication has 15 references indexed in Scilit:
- Non-cycloplegic screening for refractive errors in children with the hand-held autorefractor Retinomax: Final results and comparison with non-cycloplegic photoscreeningStrabismus, 2001
- Screening for refractive errors in children: accuracy of the hand held refractor Retinomax to screen for astigmatismBritish Journal of Ophthalmology, 1999
- Screening for abnormal levels of hyperopia in children: a non-cycloplegic method with a hand held refractorBritish Journal of Ophthalmology, 1998
- Frühzeitige Korrektion von Hyperopie und Astigmatismus bei Kindern führt zu besserer Entwicklung der SehschärfeKlinische Monatsblätter für Augenheilkunde, 1996
- Results of photorefractometric screening for amblyogenic defects in children aged 20 monthsBehavioural Brain Research, 1992
- Prediction of amblyopia and squint by means of refraction at age 1 year.British Journal of Ophthalmology, 1986
- Screening for visual defects in preschoolchildren.British Journal of Ophthalmology, 1986
- Screening for refractive errors in 6-9 month old infants by photorefraction.British Journal of Ophthalmology, 1984
- Changes in refraction between the ages of 1 and 3 1/2 years.British Journal of Ophthalmology, 1979
- Screening for refractive errors at age 1 year: a pilot study.British Journal of Ophthalmology, 1979