Purpose: To investigate characteristic spectral parameters and quality features of in-vivo 1H-MRS at 7.0 T and to measure metabolite concentrations in an intra-individual comparison to results obtained at 3.0 T and 1.5 T. Methods: Single-voxel proton MR spectra of parietal white matter were obtained at an ultra-highfield whole body MR unit operating at 7.0 T (Achieva 7.0T, Philips Medical Systems) and at clinical whole body 3.0 T and 1.5 T MR systems (Achieva 3.0T and Intera 1.5T, Philips Medical Systems). Transmit/receive quadrature head coils were used at all field strengths, and the same 3 healthy volunteers were investigated at 7.0 T, 3.0 T, and 1.5 T. PRESS-localized spectra were acquired with TR/TE 2000/47ms and water suppression by dual inversion prepulses, and VOI size (4–8cm3) and localization were held identical at the three field strengths in each volunteer. Absolute concentrations were determined by reference to the internal water signal in unsuppressed 1H spectra. Postprocessing was performed by time-domain quantification using the MRUI software package. Intra-individual comparison of spectral resolution, SNR, metabolite ratios and absolute concentrations was made at all field strengths. Results: Although magnetic field homogeneity was improved by higher order B0 shimming at 7.0 T and 3.0 T, absolute spectral linewidths (LW) increased at higher field strength due to susceptibility broadening. Nevertheless, taking into account the more than doubled chemical shift dispersion, a net increase in relative line separation Δν/LW (and thus in spectral resolution) of about 40% was achieved at 7.0 T in comparison to 3.0 T. Line broadening also caused that SNR was only about 20% higher at 7.0 T in relation to 3.0 T when calculated from peak amplitudes. However, when considering spectral line integrals, the increase at 7.0 T was close to the predicted linear field dependence. Measurements of metabolite ratios and concentrations at the 3 different field strengths yielded consistent results for choline and creatine with intra-individual variations Conclusion: The results from initial experience with in-vivo 1H-MRS at 7.0 T show that at present the observed gain in spectral resolution and SNR at ultra-high magnetic fields is limited by susceptibility-induced line broadening. The required increase in rf bandwidth and pulse duration at 7.0 T prolongs the echo time in PRESS-localized acquisitions, leading to signal dephasing for spin-coupled multiplets. Although shorter TE can be achieved with STEAM localization, the strong inherent signal loss of this technique may counter-balance the highfield SNR gain. To reduce inhomogeneity line broadening caused by intra-voxel susceptibility, application of spectroscopic imaging could be promising.