Impairing Factors in Remote-PPG Pulse Transit Time Measurements on the Face
Authors:
- Andreia MocoAMAndreia Moco
- Sander StuijkSSSander Stuijk
- Mark van GastelMGMark van Gastel
- Gerard de HaanGHGerard de Haan
- 1 June 2018
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 1439-14398
Abstract
The time it takes for a pulse wave to propagate between two arterial sites-i.e., the pulse transit time (PTT)- has received considerable attention as a marker of aortic stiffness and as a surrogate for blood pressure. However, obtrusiveness and manual intervention requirements render conventional PTT measurement methods inappropriate for ubiquitous monitoring. In this regard, high-speed camera systems are interesting alternatives. Recognizably, a technical breakthrough would be estimating PTT with a relatively inexpensive RGB camera pointed at the face only. A simple means to do this is determining the phase shift (PS) between photoplethysmographic (PPG) signals extracted at collocated skin pixels. In this paper, we show that the validity of this approach is threatened by skin variability. We analysed simultaneous video recordings of the neck and face in 21 subjects (ages, 33 ± 11 yrs). These were used to extract PPG signals at the face and skin motion (sMOT) signals at the vicinity of the carotid artery. Using sMOT as reference signal, we show that the pressure wave undergoes delay and frequency leakage as it propagates across the arterial tree; the extent of propagation distortion is subject-dependent and place PPG-based estimations at a disadvantage in comparison with PTTs measured at the arterial level. Awareness is further raised for the site-dependency of PS outcomes by the provision of facial PPG-phase maps and collocated PPG signals. Lastly, impairments due to waveform dissimilarity are demonstrated under exercise-induced PTT changes. In conclusion, PS is unsuitable for PTT measurements at the face.Keywords
This publication has 34 references indexed in Scilit:
- Robust Pulse Rate From Chrominance-Based rPPGIEEE Transactions on Biomedical Engineering, 2013
- Eulerian video magnification for revealing subtle changes in the worldACM Transactions on Graphics, 2012
- Robust smoothing of gridded data in one and higher dimensions with missing valuesComputational Statistics & Data Analysis, 2010
- Continuous and Noninvasive Blood Pressure Measurement: A Novel Modeling Methodology of the Relationship Between Blood Pressure and Pulse Wave VelocityAnnals of Biomedical Engineering, 2009
- Determining carotid artery pressure from scaled diameter waveforms: comparison and validation of calibration techniques in 2026 subjectsPhysiological Measurement, 2008
- Optical Vibrocardiography: A Novel Tool for the Optical Monitoring of Cardiac ActivityAnnals of Biomedical Engineering, 2006
- Pulse wave analysis and pulse wave velocityJournal Of Hypertension, 2003
- Pulse Transit Time as an Indicator of Arterial Blood PressurePsychophysiology, 1981
- Pulse Wave Velocity as a Measure of Blood Pressure ChangePsychophysiology, 1976
- The static elastic properties of the arterial wallThe Journal of Physiology, 1961