Tracheal Ulceration and Obstruction Associated with Flexible Bivona Tracheostomy Tubes
Authors:
Open Access
- 1 August 2006
- journal article
- case report
- Published by SAGE Publications in Anaesthesia and Intensive Care
- Vol. 34 (4) , 495-497
Abstract
An adjustable length tracheostomy tube, Bivona™, which assumes a curve on insertion, has a tendency to straighten itself in situ. The straightening force, measured electronically, was maximal (0.21 Newton) when a tube was bent 90 degrees. We observed particular clinical disadvantages of these tubes—that of tracheal ulceration (1 case), distortion of soft tracheal graft tissue (1 case) and airway obstruction when the tip embedded into the tracheal wall (1 case).Keywords
This publication has 1 reference indexed in Scilit:
- An unusual complication of a Bivona ® Hyperflex tracheostomy tubeAnaesthesia, 2005