Investigation on the reduction of electron contamination with a 6‐MV x‐ray beam
Authors:
- 1 January 1989
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 16 (1) , 123-125
Abstract
Experimental investigations have been carried out on the reduction of electron contamination of a 6‐MV x‐ray beam of Clinac model 1800 for square field sizes 5×5 to 30×30 cm2 in steps of 5 cm and for rectangular field sizes 19×7 and 7×19 cm2. The electron contamination of both the open beam and the beam with the tray can be effectively reduced by placing a lead foil filter immediately below the blocking tray. Measurements at 100‐cm source‐skin distance with filter in place showed a reduction in dose in the buildup region and also a displacement of the location of Dmax to greater depths, even for small field sizes such as 10×10 cm2.Keywords