Fetal Radiation Dose During Gestation Estimated on an Anthropomorphic Phantom for Three Generations of CT Scanners
Authors:
- Anthony G. GiletAGAnthony G. Gilet
- Jared M. DunkinJDJared M. Dunkin
- Thomas J. FernandezTFThomas J. Fernandez
- Terry M. ButtonTBTerry M. Button
- Nancy E. BudorickNBNancy E. Budorick
- 1 May 2011
- journal article
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 196 (5) , 1133-1137
Abstract
OBJECTIVE. The purpose of this study is to determine fetal dose during four different stages of pregnancy for both pulmonary CT angiogram and abdominal and pelvic CT examination on 4-, 16-, and 64-MDCT scanners measured in an anthropomorphic phantom simulating a pregnant patient. MATERIALS AND METHODS. Pulmonary angiograms and abdominal and pelvic studies were performed on a phantom on 4-, 16-, and 64-MDCT scanners. Fetal positioning and mean fetal depth were determined using data from ultrasound examinations of a large cohort of patients. Scans were performed for early pregnancy and for 10, 18, and 36 weeks. Gestational age, fetal dose, and entrance skin exposure were measured. RESULTS. When constant parameters were used for pulmonary CT angiograms, the fetal radiation dose was not significantly associated with gestational age. For abdominal examinations, the 64-MDCT scanner imparted a 20% higher dose during the third trimester than did the other scanners. When scanning parameters were kept constant between machines, gestational age and fetal dose were not significantly different. However, when the manufacturer-recommended protocols for pregnant patients were used, the dose was significantly higher in the third trimester on the 64-MDCT scanner. CONCLUSION. The 64-MDCT scanner is the most dose-efficient machine when the fetus is outside the direct scan volume, as in the case of pulmonary angiograms. For abdominal examinations, the 64-MDCT scanner imparted the highest fetal dose. This finding is attributable to the increased tube current used to penetrate the larger amount of soft tissue in late pregnancy. Abdominal shielding may reduce fetal dose without affecting diagnostic ability.Keywords
This publication has 17 references indexed in Scilit:
- Utilization of Imaging in Pregnant Patients: 10-year Review of 5270 Examinations in 3285 Patients—1997–2006Radiology, 2009
- Childhood leukaemia following medical diagnostic exposure to ionizing radiation in utero or after birthRadiation Protection Dosimetry, 2008
- Fetal radiation dose from CT pulmonary angiography in late pregnancy: a phantom studyBritish Journal of Radiology, 2008
- Radiation Exposure and Pregnancy: When Should We Be Concerned?RadioGraphics, 2007
- Internal Barium Shielding to Minimize Fetal Irradiation in Spiral Chest CT: A Phantom Simulation ExperimentRadiology, 2006
- Strategies for CT Radiation Dose OptimizationRadiology, 2004
- Monte Carlo mixture model of lifetime cancer incidence risk from radiation exposure on shuttle and international space stationMutation Research - Fundamental and Molecular Mechanisms of Mutagenesis, 1999
- Fetal position and size data for dose estimation.British Journal of Radiology, 1999
- Radiation Effective Doses to Patients Undergoing Abdominal CT ExaminationsRadiology, 1999
- Radiation exposure and the pregnant patient.American Journal of Roentgenology, 1996