Abstract
The accuracy of IOL calculation using the SRK II formula was studied in 515 cataract extractions with posterior chamber IOLs. All excessively myopic patients (8 patients) and those where we had predicted an emmetropic postoperative result, from a consecutive series of 994 patients, were included. Preoperatively the patients were divided into different groups according to their refractive status and the mean postoperative refraction was calculated in each group. The mean postoperative refraction increased almost linearly with increasing myopic status. The emmetropic group achieved a mean postoperative refraction of ‐0.6 D, whilst in the most myopic group mean refraction was ‐1.8 D. We believe that the SRK II formula is inaccurate for myopic eyes, and that new formulas are needed, taking into account all those factors that make up the dioptric power of an eye.Keywords
This publication has 14 references indexed in Scilit:
- The refractive error after implantation of a posterior chamber intraocular lensActa Ophthalmologica, 1994
- The Hoffer Q formula: A comparison of theoretic and regression formulasJournal of Cataract & Refractive Surgery, 1993
- Sources of error in intraocular lens power calculationJournal of Cataract & Refractive Surgery, 1992
- Accuracy of the newer generation intraocular lens power calculation formulas in long and short eyesJournal of Cataract & Refractive Surgery, 1991
- Clinical variability in keratometry, ultrasound biometry measurements, and emmetropic intraocular lens power calculationJournal of Cataract & Refractive Surgery, 1991
- An analysis of the accuracy of prediction of intraocular implant power in the myopic patientEye, 1991
- Effects of intraocular lens position errors on postoperative refractive errorJournal of Cataract & Refractive Surgery, 1990
- Theoretical versus SRK I and SRK II calculation of intraocular lens powerJournal of Cataract & Refractive Surgery, 1990
- Comparison of the SRK II™ formula and other second generation formulasJournal of Cataract & Refractive Surgery, 1988
- Improving the Predictability of Intraocular Lens Power CalculationsArchives of Ophthalmology (1950), 1986