Radiation Dose Reduction in Cancer Imaging with New-Model CT Scanners: A Quality of Care Evaluation
Authors:
Open Access
Abstract
Background/Objectives: The primary aim of this study was to evaluate whether the replacement of roughly one-decade-old computed tomography (CT) scanners with new-model CT scanners were associated with an additional reduction in the radiation dose delivered to oncological patients, in a radiological setting where the optimization of protocols had already reached very low radiation doses. An exploratory secondary objective was to evaluate the potential differences in the objective image quality between the CT scans obtained before and after the installation of the new-generation CT scanners. Methods: Chest and abdominal CT examinations conducted for oncologic purposes were retrospectively selected from two time periods—prior to scanner replacement (2022) and following an upgrade (2024)—after five CT systems in our radiology department were replaced. We extracted and compared the CT dose index (CTDI) and dose length product (DLP) for each CT phase. For the objective image quality evaluation, we calculated the signal-to-noise ratio (SNR) and the contrast-to-noise ratio (CNR) at the center of the liver and the aorta. An appropriate statistical analysis was performed and a p-value < 0.05 was considered significant. Results: We included 14,601 CT acquisitions, of which 9013 (61.7%) were performed before and 5588 (38.3%) after the replacement of the CT scanners. There were significantly lower values for the CTDI and DLP with the new CT scanners compared to the old ones. The CTDI with the new CT scanners was significantly lower in all phases (p-value = 0.002 for unenhanced phase, and p < 0.001 for arterial, portal venous, and delayed phases). The DLP using the new CT scanners was significantly lower in the arterial, portal venous, and delayed phases (p < 0.001), and it was not significantly different in the unenhanced phase (p = 0.36). There was no significant difference in the SNR at the liver level (p = 0.72) or at the aorta level (p = 0.51). There was no significant difference in the CNR at the liver level (p = 0.24), whereas the CNR was higher with the new CT scanners at the aorta level (p = 0.03). Conclusions: The transition to new-model CT scanners resulted in a significant reduction in the radiation dose delivered by chest and abdomen CT scans, without compromising the objective image quality.Keywords
This publication has 46 references indexed in Scilit:
- CT Scans Linked to More Cancer Cases Than Previously EstimatedJAMA, 2025
- Ovarian cancer staging and follow-up: updated guidelines from the European Society of Urogenital Radiology female pelvic imaging working groupEuropean Radiology, 2025
- Body CT examinations in oncologic patients: the impact of subspecialty radiology on radiation exposure in the clinical practice. A quality care studyLa radiologia medica, 2024
- Effect of tin spectral filtration on organ and effective dose in CT colonography and CT lung cancer screeningMedical Physics, 2023
- Vertical Off-Centering in Reduced Dose Chest-CT: Impact on Effective Dose and Image Noise ValuesAcademic Radiology, 2020
- Cumulative radiation dose from multiple CT examinations: stronger justification, fewer repeats, or dose reduction technology needed?European Radiology, 2020
- Prospective evaluation of ultra-low-dose contrast-enhanced 100-kV abdominal computed tomography with tin filter: effect on radiation dose reduction and image quality with a third-generation dual-source CT systemEuropean Radiology, 2018
- Helical CT with variable target noise levels for dose reduction in chest, abdomen and pelvis CTEuropean Radiology, 2018
- Vertical off‐centering affects organ dose in chest CT: Evidence from Monte Carlo simulations in anthropomorphic phantomsMedical Physics, 2017
- Estimation of Observer Performance for Reduced Radiation Dose Levels in CTAcademic Radiology, 2017