The reversed umbrella: displacement of a Contour device into an MCA aneurysm 18 months after treatment
- 20 March 2023
- journal article
- editorial
- Published by BMJ in Journal of NeuroInterventional Surgery
- Vol. 16 (2) , 213-215
Abstract
Summary: The Contour is a promising new device designed to treat large-neck intra-cranial aneurysms. We describe a first case of Contour device displacement 18 months after initial treatment.A patient with a 10 mm unruptured right middle cerebral artery bifurcation aneurysm was treated with a 9 mm Contour. The device was correctly positioned at the neck during treatment and at the 6 month angiography follow-up. At 18 months follow-up we noticed a full displacement of the device into the aneurysm dome. The Contour had a reversed shape and the aneurysm was still fully opacified. No neurological event occurred during the whole follow-up. Contour may be a promising tool but needs to be assessed over a long-term period.Keywords
This publication has 6 references indexed in Scilit:
- Contour-Assisted coiling with jailed microcatheter May result in better occlusion (CoCoJaMBO) in wide-necked intracranial aneurysms: Proof of principle and immediate angiographic resultsInterventional Neuroradiology, 2022
- The Safety and Effectiveness of the Contour Neurovascular System (Contour) for the Treatment of Bifurcation Aneurysms: The CERUS StudyNeurosurgery, 2021
- Woven EndoBridge device shape modification can be mitigated with an appropriate oversizing strategy: a VasoCT based studyJournal of NeuroInterventional Surgery, 2021
- The Contour—Early Human Experience of a Novel Aneurysm Occlusion DeviceClinical Neuroradiology, 2020
- Endovascular treatment of wide-necked intracranial aneurysms using the novel Contour Neurovascular System: a single-center safety and feasibility studyJournal of NeuroInterventional Surgery, 2020
- Remnants and Recurrences After the Use of the WEB Intrasaccular Device in Large-Neck Bifurcation AneurysmsNeurosurgery, 2015