Abstract
We studied temperature stability in 22 infants of birthweight less than 1500 g in the first four days of life. Infants were nursed in incubators using either air mode control or skin temperature servo control. Data were collected continuously using a computer linked monitoring system. Skin temperature control resulted in a less stable thermal environment than air mode control. Increased thermal stability in the incubator on air mode control may well be beneficial, particularly to sick, very low birthweight infants.Keywords
This publication has 14 references indexed in Scilit:
- New standards for neutral thermal environment of healthy very low birthweight infants in week one of life.Archives of Disease in Childhood, 1984
- Air versus skin temperature servocontrol of infant incubatorsThe Journal of Pediatrics, 1983
- Incubation of very immature infants.Archives of Disease in Childhood, 1983
- Temperature Regulation in Sick Infants1983
- Effects of Low Humidity on Small Premature Infants in Servocontrol IncubatorsNeonatology, 1975
- Preliminary studies on the thermal environment of low-birth-weight infantsThe Journal of Pediatrics, 1973
- The Optimum Thermal Environment for Naked BabiesArchives of Disease in Childhood, 1970
- Apnea in Premature Infants and Incubator-Air-Temperature ChangesNew England Journal of Medicine, 1970
- Relationship of Thermal Environment and Caloric Intake to Growth and Resting Metabolism in the Late Neonatal PeriodNeonatology, 1969
- EFFECT OF THE THERMAL ENVIRONMENT ON COLD RESISTANCE AND GROWTH OF SMALL INFANTS AFTER THE FIRST WEEK OF LIFEPediatrics, 1968