Validation of the Oncomine™ focus panel for next-generation sequencing of clinical tumour samples
Open Access
- 13 August 2018
- journal article
- research article
- Published by Springer Nature in Virchows Archiv
- Vol. 473 (4) , 489-503
Abstract
The clinical utility of next-generation sequencing (NGS) for a diverse range of targets is expanding, increasing the need for multiplexed analysis of both DNA and RNA. However, translation into daily use requires a rigorous and comprehensive validation strategy. The aim of this clinical validation was to assess the performance of the Ion Torrent Personal Genome Machine (IonPGM™) and validate the Oncomine™ Focus DNA and RNA Fusion panels for clinical application in solid tumour testing of formalin-fixed, paraffin-embedded (FFPE) tissue. Using a mixture of routine FFPE and reference material across a variety of tissue and specimen types, we sequenced 86 and 31 samples on the Oncomine™ Focus DNA and RNA Fusion assays, respectively. This validation considered a number of parameters including the clinical robustness of the bioinformatics pipeline for variant detection and interpretation. The Oncomine™ Focus DNA assay had a sample and variant-based sensitivity of 99.1 and 97.1%, respectively, and an assay specificity of 100%. The Oncomine™ Focus Fusion panel had a good sensitivity and specificity based upon the samples assessed, however requires further validation to confirm findings due to limited sample numbers. We observed a good sequencing performance based upon amplicon, gene (hotspot variants within gene) and sample specific analysis with 92% of clinical samples obtaining an average amplicon coverage above 500X. Detection of some indels was challenging for the routine IonReporter™ workflow; however, the addition of NextGENe® software improved indel identification demonstrating the importance of both bench and bioinformatic validation. With an increasing number of clinically actionable targets requiring a variety of methodologies, NGS provides a cost-effective and time-saving methodology to assess multiple targets across different modalities. We suggest the use of multiple analysis software to ensure identification of clinically applicable variants.Keywords
Funding Information
- University of St. Andrews
This publication has 18 references indexed in Scilit:
- Guidelines for Validation of Next-Generation Sequencing–Based Oncology PanelsThe Journal of Molecular Diagnostics, 2017
- Clinical Validation of Targeted Next Generation Sequencing for Colon and Lung CancersPLOS ONE, 2015
- Sequence Artifacts in DNA from Formalin-Fixed Tissues: Causes and Strategies for MinimizationClinical Chemistry, 2015
- Evaluation and optimisation of indel detection workflows for ion torrent sequencing of the BRCA1 and BRCA2 genesBMC Genomics, 2014
- DNA Qualification Workflow for Next Generation Sequencing of Histopathological SamplesPLOS ONE, 2013
- Shining a Light on Dark Sequencing: Characterising Errors in Ion Torrent PGM DataPLOS Computational Biology, 2013
- Performance comparison of benchtop high-throughput sequencing platformsNature Biotechnology, 2012
- Integrative Genomics Viewer (IGV): high-performance genomics data visualization and explorationBriefings in Bioinformatics, 2012
- Effect of Fixatives and Tissue Processing on the Content and Integrity of Nucleic AcidsThe American Journal of Pathology, 2002
- PCR Amplification from Paraffin-Embedded Tissues: Effects of Fixative and Fixation TimeAmerican Journal of Clinical Pathology, 1991