The Determination of IgG Subclass Concentrations in Serum by Enzyme-Linked Immunosorbent Assay: Establishment of Age-Related Reference Ranges for Cord Blood Samples, Children Aged 5–13 Years and Adults
Authors:
Open Access
- 1 May 1994
- journal article
- research article
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Vol. 31 (3) , 245-248
Abstract
We describe a simple enzyme-linked immunosorbent assay for the measurement of immunoglobulin G (IgG) subclasses in serum. The microtitre plate is coated directly with diluted samples, standards or controls, blocked with phosphate-buffered saline containing 1% Tween to prevent non-specific protein binding to the plate, and each subclass assigned with specific mouse monoclonal antibody. The specific antibody is detected with alkaline-phosphatase linked anti-globulin and the colour development of the paranitrophenyl phosphate indicator reagent is measured at 405 nm until the highest concentration of standard gives an absorbance in excess of 1·4. Coefficients of variation measured at three concentration levels were less than 7% for within-assay variation and less than 10% for between-assay variation. Regression analysis of total IgG with the sum of the measured IgG subclasses gave a correlation coefficient of 0·932 ( P < 0·001). The assay has been used to establish age-related reference ranges for serum IgG subclasses; these are found to be slightly different from those established for radioimmunodiffusion methods. Our study does not confirm the previously reported absence of IgG 4 in normal individuals.Keywords
This publication has 10 references indexed in Scilit:
- Chicken antibodies: a tool to avoid interference by complement activation in ELISAJournal of Immunological Methods, 1992
- Immunodeficiency presenting as hypergammaglobulinaemia with IgG2 subclass deficiencyThe Lancet, 1992
- Abnormal serum IgG subclass pattern in children with Down's syndrome.Archives of Disease in Childhood, 1992
- Interpretation of IgG subclass values: A comparison of two assaysThe Journal of Pediatrics, 1990
- Selective IgG subclass deficiency: quantification and clinical relevanceClinical and Experimental Immunology, 1990
- Serum IgG subclass levels in patients with primary immunodeficiency syndromes or abnormal susceptibility to infectionsClinical Immunology and Immunopathology, 1989
- Monoclonal antibody-based solid-phase immunoenzymometric assays for quantifying human immunoglobulin G and its subclasses in serum.Clinical Chemistry, 1985
- Impaired Lung Function in Patients with IgA Deficiency and Low Levels of IgG2 or IgG3New England Journal of Medicine, 1985
- IgG Subclasses in Selective IgA DeficiencyNew England Journal of Medicine, 1981
- Selective Gamma-G Globulin Deficiencies in Patients with Recurrent Pyogenic InfectionsNew England Journal of Medicine, 1970