A multichannel, real‐time MRI RF power monitor for independent SAR determination
Authors:
- 11 April 2012
- journal article
- Published by Wiley in Medical Physics
- Vol. 39 (5) , 2334-2341
Abstract
Purpose: Accurate measurements of the RF power delivered during clinical MRI are essential for safety and regulatory compliance, avoiding inappropriate restrictions on clinical MRIsequences, and for testing the MRIsafety of peripheral and interventional devices at known RF exposure levels. The goal is to make independent RF power measurements to test the accuracy of scanner-reported specific absorption rate (SAR) over the extraordinary range of operating conditions routinely encountered in MRI. Methods: A six channel, high dynamic range, real-time power profiling system was designed and built for monitoring power delivery during MRI up to 440 MHz. The system was calibrated and used in two 3 T scanners to measure power applied to human subjects during MRI scans. The results were compared with the scanner-reported SAR. Results: The new power measurement system has highly linear performance over a 90 dB dynamic range and a wide range of MRI duty cycles. It has about 0.1 dB insertion loss that does not interfere with scanner operation. The measurements of whole-body SAR in volunteers showed that scanner-reported SAR was significantly overestimated by up to about 2.2 fold. Conclusions: The new power monitor system can accurately and independently measure RF power deposition over the wide range of conditions routinely encountered during MRI. Scanner-reported SAR values are not appropriate for setting exposure limits during device or pulse sequencetesting.Keywords
This publication has 35 references indexed in Scilit:
- Toward individualized SAR models and in vivo validationMagnetic Resonance in Medicine, 2011
- Calculation of radiofrequency electromagnetic fields and their effects in MRI of human subjectsMagnetic Resonance in Medicine, 2011
- Simultaneous B0‐ and B1+‐Map acquisition for fast localized shim, frequency, and RF power determination in the heart at 3 TMagnetic Resonance in Medicine, 2010
- Experimental and numerical assessment of MRI‐induced temperature change and SAR distributions in phantoms and in vivoMagnetic Resonance in Medicine, 2009
- Noise figure limits for circular loop MR coilsMagnetic Resonance in Medicine, 2009
- Turning Up the Heat on MRIJournal of the American College of Radiology, 2008
- A prototype RF dosimeter for independent measurement of the average specific absorption rate (SAR) during MRIJournal of Magnetic Resonance Imaging, 2007
- Calculations ofB 1 distribution, specific energy absorption rate, and intrinsic signal-to-noise ratio for a body-size birdcage coil loaded with different human subjects at 64 and 128 MHzApplied Magnetic Resonance, 2005
- Radiofrequency map of an NMR coil by imagingMagnetic Resonance Imaging, 1993
- Estimating radiofrequency power deposition in body NMR imagingMagnetic Resonance in Medicine, 1985