Optimal Expiratory Volume Profile in Tidal Liquid Ventilation under Steady State Conditions, Based on a Symmetrical Lung Model
Authors:
- 1 January 2009
- journal article
- respiratory support
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 55 (1) , 63-72
Abstract
Liquid-assisted ventilation (LAV) using perfluorochemicals (PFC) offers clear theoretical advantages over gas ventilation. During tidal liquid ventilation (TLV) the residual capacity of the lungs is filled with PFC and a liquid ventilator is necessary to inhale and exhale the tidal volume of PFC. However, during the expiration phase, a flow limitation (choked flow) can be observed, which compromises minute ventilation and consequently the gas exchange. The hypothesis of the presented works is that the choked flow can be avoided by profiling the expiratory volume. To validate this concept, an elastic symmetrical lung numerical model, used to characterize forced expiration in gas ventilation, was transposed to TLV. The parameters of the developed numerical model were fitted from experimental data obtained on a newborn lamb. The results obtained demonstrate that general observations made with gas ventilation still hold, however, in TLV: flow limitation in the central airways is the result of a coupling between viscous pressure losses and airway compliance, and the flow limiting segment is located in the central airways. Using the model results, an optimal theoretical expiratory profile seems to be exponential as first approximation, and its time constant is dependent on the chocked flow mechanism and not on the product of resistance by compliance. This optimal profile is used to compute the maximal minute ventilation allowable with an acceptable risk of collapse. Also, the sensitivity of minute ventilation to different parameter variations were analyzed and practical recommendations are proposed.Keywords
This publication has 43 references indexed in Scilit:
- Influence of Partial Liquid Ventilation on Bacterial Growth and Alveolar Expansion in Newborn Rabbits with Group B-Streptococcal PneumoniaPediatric Research, 2003
- Long-Term Tidal Liquid Ventilation in Premature Lambs: Physiologic, Biochemical and Histological CorrelatesNeonatology, 2003
- Perfluorocarbons attenuate oxidative lung damagePediatric Pulmonology, 2003
- Prospective, Randomized, Controlled Pilot Study of Partial Liquid Ventilation in Adult Acute Respiratory Distress SyndromeAmerican Journal of Respiratory and Critical Care Medicine, 2002
- Partial liquid ventilation reduces pulmonary neutrophil accumulation in an experimental model of systemic endotoxemia and acute lung injuryCritical Care Medicine, 1998
- Evaluation of gas exchange, pulmonary compliance, and lung injury during total and partial liquid ventilation in the acute respiratory distress syndromeCritical Care Medicine, 1996
- Liquid assisted ventilation: An alternative ventilatory strategy for acute meconium aspiration injuryPediatric Pulmonology, 1996
- Improvement of Gas Exchange, Pulmonary Function, and Lung Injury With Partial Liquid VentilationChest, 1995
- Lung lavage with oxygenated perfluorochemical liquid in acute lung injuryCritical Care Medicine, 1993
- Survival of Mammals Breathing Organic Liquids Equilibrated with Oxygen at Atmospheric PressureScience, 1966