The Validity of Bispectral Index Values from a Dislocated Sensor: A Comparison with Values from a Sensor Located in the Commercially Recommended Position
Authors:
- Masahiko KawaguchiMKMasahiko Kawaguchi
- Naoko KuritaNKNaoko Kurita
- Satoki InoueSISatoki Inoue
- Hitoshi FuruyaHFHitoshi Furuya
- 1 April 2007
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 104 (4) , 857-859
Abstract
The influence of sensor dislocation on bispectral index (BIS) values is not clear. We compared the BIS values obtained from dislocated sensors with those from the commercially recommended positions. We used two BIS sensors for each patient receiving propofol-based anesthesia; one in the recommended position and one positioned around the lateral corner of the right eye. Bland and Altman analysis revealed better agreement of two BIS values when the values during induction of and emergence from anesthesia were excluded. The results indicate that during induction of and emergence from general anesthesia, a dislocated BIS sensor may produce questionable BIS values.Keywords
This publication has 7 references indexed in Scilit:
- Bispectral Index monitoring correlates with sedation scales in brain-injured patients*Critical Care Medicine, 2004
- An Unusual Site for BIS MonitoringAnesthesia & Analgesia, 2004
- A Comparison of Frontal and Occipital Bispectral Index Values Obtained During Neurosurgical ProceduresAnesthesia & Analgesia, 2004
- Mild Core Hypothermia and Anesthetic Requirement for Loss of Responsiveness During Propofol Anesthesia for CraniotomyAnesthesia & Analgesia, 2002
- Monitoring Depth of AnesthesiaAnesthesiology, 2000
- Bispectral Analysis Measures Sedation and Memory Effects of Propofol, Midazolam, Isoflurane, and Alfentanil in Healthy VolunteersAnesthesiology, 1997
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986