Characterization of the a-Si EPID in the unity MR-linac for dosimetric applications
- 1 January 2018
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 63 (2) , 025006
Abstract
Electronic portal imaging devices (EPIDs) are frequently used in external beam radiation therapy for dose verification purposes. The aim of this study was to investigate the dose-response characteristics of the EPID in the Unity MR-linac (Elekta AB, Stockholm, Sweden) relevant for dosimetric applications under clinical conditions. EPID images and ionization chamber (IC) measurements were used to study the effects of the magnetic field, the scatter generated in the MR housing reaching the EPID, and inhomogeneous attenuation from the MR housing. Dose linearity and dose rate dependencies were also determined. The magnetic field strength at EPID level did not exceed 10 mT, and dose linearity and dose rate dependencies proved to be comparable to that on a conventional linac. Profiles of fields, delivered with and without the magnetic field, were indistinguishable. The EPID center had an offset of 5.6 cm in the longitudinal direction, compared to the beam central axis, meaning that large fields in this direction will partially fall outside the detector area and not be suitable for verification. Beam attenuation by the MRI scanner and the table is gantry angle dependent, presenting a minimum attenuation of 67% relative to the 90° measurement. Repeatability, observed over two months, was within 0.5% (1 SD). In order to use the EPID for dosimetric applications in the MR-linac, challenges related to the EPID position, scatter from the MR housing, and the inhomogeneous, gantry angle-dependent attenuation of the beam will need to be solved.Keywords
Funding Information
- Elekta
This publication has 28 references indexed in Scilit:
- Gantry-angle resolved VMAT pretreatment verification using EPID image predictionMedical Physics, 2013
- Dose response of selected ion chambers in applied homogeneous transverse and longitudinal magnetic fieldsMedical Physics, 2013
- In aqua vivo EPID dosimetryMedical Physics, 2011
- A literature review of electronic portal imaging for radiotherapy dosimetryRadiotherapy and Oncology, 2008
- Dosimetric pre-treatment verification of IMRT using an EPID; clinical experienceRadiotherapy and Oncology, 2006
- Comparison of ghosting effects for three commercial EPIDsMedical Physics, 2006
- Dose‐response characteristics of an amorphous silicon EPIDMedical Physics, 2005
- The long‐term stability of amorphous silicon flat panel imaging devices for dosimetry purposesMedical Physics, 2004
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapyMedical Physics, 2003
- An investigation of a new amorphous silicon electronic portal imaging device for transit dosimetryMedical Physics, 2002