Outcomes following pacemaker implantation after transcatheter aortic valve implantation with CoreValve® devices: Results from the FRANCE 2 Registry

Authors:

Abstract
Aims: Permanent pacemaker (PPM) implantation following high‐degree atrioventricular block is frequently required after transcatheter aortic valve implantation (TAVI) using CoreValve®. Recent improvement of the delivery system (CoreValve Accutrak®) aimed to ease delivery and reduce the PPM rate. Our study evaluated the incidences of PPM implantation following use of CoreValve® or CoreValve Accutrak®and the clinical outcome of these patients.Methods and results: A total of 883 patients (82 ± 7 years; 41.3% female) with severe symptomatic aortic stenosis and self‐expanding bioprosthesis implantation were included between January 2010 and October 2011 in 29 centers from the FRANCE 2 Registry. Follow‐up data were available in 833 patients. CoreValve® and CoreValve Accutrak® were used in 343 (41.2%) and 490 (58.8%) patients, respectively. During a mean follow‐up of 242 ± 179 days, all‐cause mortality was similar in patients with versus without PPM implantation (16.3 vs. 16.9%, P = 0.832).There was no significant difference in the PPM incidence in CoreValve® and CoreValve Accutrak®patients (30.4% vs. 27.5%, P = 0.846).Conclusion: PPM implantation remained frequent after TAVI using CoreValve Accutrak®. All‐cause mortality was similar in patients with or without PPM implantation. The new device failed to show a significant decrease in PPM implantation incidence after TAVI.

This publication has 23 references indexed in Scilit: