Ventricular capture by anodal pacemaker stimulation
Open Access
- 23 March 2006
- journal article
- case report
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 8 (5) , 385-387
Abstract
This report describes the case of an 86-year-old male with syncopal paroxysmal 2:1 atrioventricular block and a single chamber VVI pacemaker programmed to bipolar sensing and unipolar pacing. After recurrence of syncope, a complete loss of ventricular capture with regular ventricular sensing was observed on ECG; fluoroscopic examination suggested perforation of the right ventricle by the helix of the implanted screw-in lead. Reprogramming the pacemaker to bipolar pacing/sensing resulted in regular ventricular capture and sensing, suggesting effective anodal stimulation from the ring electrode permitting complete non-invasive palliation.Keywords
This publication has 6 references indexed in Scilit:
- Mechanism of Anode Break Stimulation in the HeartBiophysical Journal, 1998
- Virtual electrode effects in myocardial fibersBiophysical Journal, 1994
- Virtual cathode effects during stimulation of cardiac muscle. Two-dimensional in vivo experiments.Circulation Research, 1991
- Effects of unipolar cathodal and bipolar stimulation on vulnerability of ischemic ventricles to fibrillationThe American Journal of Cardiology, 1975
- Direct Current Make and Break Thresholds for Pacemaker Electrodes on the Canine VentricleCirculation Research, 1970
- Anodal Excitation of Cardiac MuscleAmerican Journal of Physiology-Legacy Content, 1957