Reproducibility of electrical impedance tomographic spectroscopy (EITS) parametric images of neonatal lungs
Authors:
- 1 November 1996
- journal article
- Published by IOP Publishing in Physiological Measurement
- Vol. 17 (4A) , A205-A212
Abstract
The reproducibility of electrical impedance tomographic spectroscopy (EITS) images of neonatal lungs have been investigated in 11 clinically stable babies. We have used the Sheffield Mark IIIa EITS system. An average inspiration frame was generated from the data frames associated with maximum inspiration. Frequency images were reconstructed from these frames. The frequency images were analysed to locate the pixel with the maximum change in the right lung field. The change was defined as the 614 kHz measurement relative to 9.6 kHz. A pixel region of interest was centred at this point. The changes in impedance with frequency for this region of interest show good overall reproducibility between electrode applications for eight frequencies (95% limits of agreement ). This reproducibility is improved (95% limits of agreement ) by omitting the highest frequency (1.2 MHz) which is most subject to system noise. The parameters for the Cole model derived from data with the highest frequency omitted are less reproducible between electrode applications (95% limits of agreement, ). We suspect that the parametric model used may have an effect on this. The signals recorded at the highest frequency (1.2 MHz) are a major source of variability. The reproducibility results are improved by omitting this frequency from the analysis.Keywords
This publication has 6 references indexed in Scilit:
- Multifrequency and parametric EIT images of neonatal lungsPhysiological Measurement, 1995
- Measured and expected Cole parameters from electrical impedance tomographic spectroscopy images of the human thoraxPhysiological Measurement, 1995
- A fast parametric modelling algorithm with the Powell methodPhysiological Measurement, 1995
- Multi-frequency imaging and modelling of respiratory related electrical impedance changesPhysiological Measurement, 1994
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Dispersion and Absorption in Dielectrics I. Alternating Current CharacteristicsThe Journal of Chemical Physics, 1941